Stereoscopic calibration for augmented reality visualization in microscopic surgery.

PURPOSE: Middle and inner ear procedures target hearing loss, infections, and tumors of the temporal bone and lateral skull base. Despite the advances in surgical techniques, these procedures remain challenging due to limited haptic and visual feedback. Augmented reality (AR) may improve operative safety by allowing the 3D visualization of anatomical structures from preoperative computed tomography (CT) scans on real intraoperative microscope video feed. The purpose of this work was to develop a real-time CT-augmented stereo microscope system using camera calibration and electromagnetic (EM) tracking.

METHODS: A 3D printed and electromagnetically tracked calibration board was used to compute the intrinsic and extrinsic parameters of the surgical stereo microscope. These parameters were used to establish a transformation between the EM tracker coordinate system and the stereo microscope image space such that any tracked 3D point can be projected onto the left and right images of the microscope video stream. This allowed the augmentation of the microscope feed of a 3D printed temporal bone with its corresponding CT-derived virtual model. Finally, the calibration board was also used for evaluating the accuracy of the calibration.

RESULTS: We evaluated the accuracy of the system by calculating the registration error (RE) in 2D and 3D in a microsurgical laboratory setting. Our calibration workflow achieved a RE of 0.11 ± 0.06 mm in 2D and 0.98 ± 0.13 mm in 3D. In addition, we overlaid a 3D CT model on the microscope feed of a 3D resin printed model of a segmented temporal bone. The system exhibited small latency and good registration accuracy.

CONCLUSION: We present the calibration of an electromagnetically tracked surgical stereo microscope for augmented reality visualization. The calibration method achieved accuracy within a range suitable for otologic procedures. The AR process introduces enhanced visualization of the surgical field while allowing depth perception.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app