Evaluation of intraoperative neuromonitoring (IONM) data with the Mainz IONM Quality Assurance and Analysis tool.

BJS Open 2023 July 11
BACKGROUND: Intraoperative neuromonitoring is widely used in thyroid and parathyroid surgery to prevent unilateral and especially bilateral recurrent nerve paresis. Reference values for amplitude and latency for the recurrent laryngeal nerve and vagus nerve have been published. However, data quality measures that exclude errors of the underlying intraoperative neuromonitoring (IONM) data (immanent software errors, false data labelling) before statistical analysis have not yet been implemented.

METHODS: The authors developed an easy-to-use application (the Mainz IONM Quality Assurance and Analysis tool) using the programming language R. This tool allows visualization, automated and manual correction, and statistical analysis of complete raw data sets (electromyogram signals of all stimulations) from intermittent and continuous neuromonitoring in thyroid and parathyroid surgery. The Mainz IONM Quality Assurance and Analysis tool was used to evaluate IONM data generated and exported from 'C2' and 'C2 Xplore' neuromonitoring devices (inomed Medizintechnik GmbH) after surgery. For the first time, reference values for latency and amplitude were calculated based on 'cleaned' IONM data.

RESULTS: Intraoperative neuromonitoring data files of 1935 patients consecutively operated on from June 2014 to May 2020 were included. Of 1921 readable files, 34 were excluded for missing data labelling. Automated plausibility checks revealed: less than 3 per cent device errors for electromyogram signal detection; 1138 files (approximately 60 per cent) contained potential labelling errors or inconsistencies necessitating manual review; and 915 files (48.5 per cent) were indeed erroneous. Mean(s.d.) reference onset latencies for the left vagus nerve, right vagus nerve, recurrent laryngeal nerve, and external branch of the superior laryngeal nerve were 6.8(1.1), 4.2(0.8), 2.5(1.1), and 2.1(0.5) ms, respectively.

CONCLUSION: Due to high error frequencies, IONM data should undergo in-depth review and multi-step cleaning processes before analysis to standardize scientific reporting. Device software calculates latencies differently; therefore reference values are device-specific (latency) and/or set-up-specific (amplitude). Novel C2-specific reference values for latency and amplitude deviate considerably from published values.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app