Concurrent kilovoltage CBCT imaging and megavoltage beam delivery: suppression of cross-scatter with 2D antiscatter grids and grid-based scatter sampling.

Objective . The concept of using kilovoltage (kV) and megavoltage (MV) beams concurrently has potential applications in cone beam computed tomography (CBCT) guided radiation therapy, such as single breath hold scans, metal artifact reduction, and simultaneous imaging during MV treatment delivery. However, MV cross-scatter generated during MV beam delivery degrades CBCT image quality. To address this, a 2D antiscatter grid and a cross-scatter correction method were investigated in the context of high dose MV treatment delivery. Approach . A 3D printed, tungsten 2D antiscatter grid prototype was utilized in kV CBCT scans to reduce MV cross-scatter fluence during concurrent MV beam delivery. Remaining cross-scatter in projections was corrected by using the 2D grid itself as a cross-scatter intensity sampling device, referred to as grid-based scatter sampling (GSS). To test this approach, kV CBCT acquisitions were performed while delivering 6 and 10 MV beams, mimicking high dose rate treatment delivery scenarios. kV and MV beam deliveries were not synchronized to eliminate MV beam delivery interruption. MV cross-scatter suppression performance of the proposed approach was evaluated in projections and CBCT images of phantoms. Main results . 2D grid reduced the intensity of MV cross-scatter in kV projections by a factor of 3 on the average, when compared to conventional antiscatter grid. Remaining cross scatter as measured by the GSS method was within 7% of measured reference intensity values, and subsequently corrected. CBCT image quality was improved substantially during concurrent kV-MV beam delivery. Median Hounsfield Unit (HU) inaccuracy was up to 182 HU without our methods, and it was reduced to a median 6.5 HU with our 2D grid and scatter correction approach. Our methods provided a factor of 2-6 improvement in contrast-to-noise ratio. Significance . This investigation demonstrates the utility of 2D antiscatter grids and grid-based scatter sampling in suppressing MV cross-scatter. Our approach successfully minimized the effects of MV cross-scatter in concurrent kV CBCT imaging and high dose MV treatment delivery scenarios. Hence, robust MV cross-scatter suppression is potentially feasible without MV beam delivery interruption or compromising kV image acquisition rate.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app