Evaluation of the B-Smart manometer and the CompuFlo computerized injection pump technology for accurate needle-tip injection pressure measurement during peripheral nerve blockade.

BACKGROUND AND OBJECTIVES: The exact mechanism of peripheral nerve blocks causing/leading to nerve injury remains controversial. Evidence from animal experiments suggests that intrafascicular injection resulting in high injection pressure has the potential to rupture nerve fascicles and may consequently cause permanent nerve injury and neurological deficits. The B-Smart (BS) in-line manometer and the CompuFlo (CF) computerized injection pump technology are two modalities used for monitoring pressure during regional anesthesia. This study sought to explore the accuracy of these two technologies in measuring needle-tip pressures in a simulated environment.

METHODS: In seven simulated needle-syringe combinations, the BS and the CF devices were connected in series through a closed system and attached to a digital manometer at the tip of various needles. The pressures were evaluated in three trials per needle-syringe combination. Sensitivity, specificity, positive predictive value, negative predictive value, and accuracy (F1 Score) were determined for each needle type and overall.

RESULTS: For pressures ≥15 psi and ≥20 psi, respectively, the CF device demonstrated a sensitivity of 100%, 100%; specificity of 96%, 98%; positive predictive value 93%, 93%; and negative predictive value of 100%, 100%. The BS device demonstrated a sensitivity of 60%, 100%; specificity of 99%, 95%; positive predictive value of 96%, 85%; and negative predictive value of 85%, 100%. Accuracy, as measured by the F1 Score, for detecting a pressure of ≥15 psi was 0.96 for the CF and 0.74 for the BS.

CONCLUSIONS: Future research is needed to explore in-vivo performance and evaluate whether either of these devices can impact on clinical outcomes.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app