Noise reduction and image quality in ultra-high resolution computed tomography of the temporal bone using advanced modeled iterative reconstruction.
Acta Radiologica 2019 January 9
BACKGROUND: The novel advanced modeled iterative reconstruction (ADMIRE) algorithm in ultra-high-resolution (UHR) computed tomography (CT) of the temporal bone has not yet been systematically evaluated.
PURPOSE: To assess the potential of ADMIRE in temporal bone UHR CT.
MATERIAL AND METHODS: Forty-four patients who underwent UHR CT of the temporal bone using z-axis UHR protocol were retrospectively selected for analysis. Images were reconstructed using filtered back projection (FBP) and ADMIRE with multiple strength levels. Regions of interest were drawn in the posterior fossa and petrous bone. The average density (in Hounsfield units [HU]) and the image noise (standard deviation of density values) were extracted. The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were then calculated. Additionally, a subjective qualitative analysis was performed using a five-point Likert scale. The potential dose reduction was extrapolated from objective analysis and confirmed in an additional phantom study.
RESULTS: The image noise was significantly lower, and the SNR and CNR were significantly higher in ADMIRE reconstructions levels A2-A5 than in FBP ( P < 0.001, respectively). Subjective image quality was significantly higher in ADMIRE levels A2-A5 than in FBP ( P < 0.001) and it was rated best in ADMIRE level A3. Confirmed by the results from the phantom study, a dose reduction of at least 40% was feasible while maintaining image quality.
CONCLUSIONS: The ADMIRE reconstruction algorithm significantly improves image quality and reduces noise on temporal bone UHR CT scans. Thus, it allows for substantial dose reduction.
PURPOSE: To assess the potential of ADMIRE in temporal bone UHR CT.
MATERIAL AND METHODS: Forty-four patients who underwent UHR CT of the temporal bone using z-axis UHR protocol were retrospectively selected for analysis. Images were reconstructed using filtered back projection (FBP) and ADMIRE with multiple strength levels. Regions of interest were drawn in the posterior fossa and petrous bone. The average density (in Hounsfield units [HU]) and the image noise (standard deviation of density values) were extracted. The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were then calculated. Additionally, a subjective qualitative analysis was performed using a five-point Likert scale. The potential dose reduction was extrapolated from objective analysis and confirmed in an additional phantom study.
RESULTS: The image noise was significantly lower, and the SNR and CNR were significantly higher in ADMIRE reconstructions levels A2-A5 than in FBP ( P < 0.001, respectively). Subjective image quality was significantly higher in ADMIRE levels A2-A5 than in FBP ( P < 0.001) and it was rated best in ADMIRE level A3. Confirmed by the results from the phantom study, a dose reduction of at least 40% was feasible while maintaining image quality.
CONCLUSIONS: The ADMIRE reconstruction algorithm significantly improves image quality and reduces noise on temporal bone UHR CT scans. Thus, it allows for substantial dose reduction.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app