Efficacy of the Abdominal Aortic Junctional Tourniquet-Torso Plate in a Lethal Model of Noncompressible Torso Hemorrhage.
Journal of Special Operations Medicine : a Peer Reviewed Journal for SOF Medical Professionals 0 December
BACKGROUND: The Abdominal Aortic Junctional Tourniquet, when modified with an off-label, prototype, accessory pressure distribution plate (AAJT-TP), has the potential to control noncompressible torso hemorrhage in prolonged field care.
METHODS: Using a lethal, noncompressible torso hemorrhage model, 24 male Yorkshire swine (81kg-96kg) were randomly assigned into two groups (control or AAJT-TP). Anesthetized animals were instrumented and an 80% laparoscopic, left-side liver lobe transection was performed. At 10 minutes, the AAJT-TP was applied and inflated to an intraabdominal pressure of 40mmHg. At 20 minutes after application, the AAJT-TP was deflated, but the windlass was left tightened. Animals were observed for a prehospital time of 60 minutes. Animals then underwent damage control surgery at 180 minutes, followed by an intensive care unit-phase of care for an additional 240 minutes. Survival was the primary end point.
RESULTS: Compared with Hextend, survival was not significantly different in the AAJT-TP group (ρ = .564), nor was blood loss (3.3L ± 0.5L and 3.0L ± 0.5L, respectively; p = .285). There was also no difference in all physiologic parameters between groups at the end of the study or end of the prehospital phase. Three of 12 AAJT-TP animals had an inferior vena cava thrombus.
CONCLUSION: The AAJT-TP did not provide any survival benefit compared with Hextend alone in this model of noncompressible torso hemorrhage.
METHODS: Using a lethal, noncompressible torso hemorrhage model, 24 male Yorkshire swine (81kg-96kg) were randomly assigned into two groups (control or AAJT-TP). Anesthetized animals were instrumented and an 80% laparoscopic, left-side liver lobe transection was performed. At 10 minutes, the AAJT-TP was applied and inflated to an intraabdominal pressure of 40mmHg. At 20 minutes after application, the AAJT-TP was deflated, but the windlass was left tightened. Animals were observed for a prehospital time of 60 minutes. Animals then underwent damage control surgery at 180 minutes, followed by an intensive care unit-phase of care for an additional 240 minutes. Survival was the primary end point.
RESULTS: Compared with Hextend, survival was not significantly different in the AAJT-TP group (ρ = .564), nor was blood loss (3.3L ± 0.5L and 3.0L ± 0.5L, respectively; p = .285). There was also no difference in all physiologic parameters between groups at the end of the study or end of the prehospital phase. Three of 12 AAJT-TP animals had an inferior vena cava thrombus.
CONCLUSION: The AAJT-TP did not provide any survival benefit compared with Hextend alone in this model of noncompressible torso hemorrhage.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app