Re-exposure in cone beam CT of the dentomaxillofacial region: a retrospective study.
Dento Maxillo Facial Radiology 2019 January 4
METHODS:: In a retrospective cohort study CBCT images of 4986 patients from the patient database from the Department of Oral Radiology, University Medical Center of the Johannes Gutenberg University, Mainz, Germany were included and the rate of re-exposures was counted. Patients were stratified into those who received a scan with the small field-of-view CBCT or the large field-of-view CBCT. The effect of patient-related parameters as age and gender was implicated. As a further device-specific parameter, the statistical analysis included whether the selection of the field of view due to the device type had a significant influence on the occurrence of re-exposures. Furthermore, the rescans were analyzed with regard to their causes.
RESULTS:: In total, CBCT images of 82 (1.6%) patients had to be repeated. Looking at the two different devices, in 42 (1.3%) patients that received a scan with the large field-of-view CBCT and in 40 (2.3%) patients that received a scan with the small field-of-view CBCT respectively needed a retake. There was no statistically significant correlation between age and gender to retakes. For the small field-of-view-size significantly more retakes were observed than for the large one. With 46% motion artifacts were the most frequent causes for a re-exposure of the patient.
CONCLUSIONS:: Gender and age did not have an impact on the occurrence of re-exposures. Patients who received a scan with the small field-of-view CBCT were significantly more often rescanned than those with the large field-of-view CBCT.
RESULTS:: In total, CBCT images of 82 (1.6%) patients had to be repeated. Looking at the two different devices, in 42 (1.3%) patients that received a scan with the large field-of-view CBCT and in 40 (2.3%) patients that received a scan with the small field-of-view CBCT respectively needed a retake. There was no statistically significant correlation between age and gender to retakes. For the small field-of-view-size significantly more retakes were observed than for the large one. With 46% motion artifacts were the most frequent causes for a re-exposure of the patient.
CONCLUSIONS:: Gender and age did not have an impact on the occurrence of re-exposures. Patients who received a scan with the small field-of-view CBCT were significantly more often rescanned than those with the large field-of-view CBCT.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
Perioperative echocardiographic strain analysis: what anesthesiologists should know.Canadian Journal of Anaesthesia 2024 April 11
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app