Add like
Add dislike
Add to saved papers

Design and Evaluation of a Wireless Electrocardiogram Monitor in an Operating Room: A Pilot Study.

Anesthesia and Analgesia 2018 December 12
BACKGROUND: Wired electrocardiogram monitors are an important component of current perioperative monitoring. Wireless monitoring units could help reduce the number of cables attached to patients and thus improve anesthesia ergonomics and patient management. However, there is concern that electromagnetic interference generated by electrosurgical units may prevent effective wireless signals in the operating room. To evaluate the extent of this problem, we developed a Bluetooth electrocardiogram prototype monitor and compared its electrocardiogram traces to those captured with a standard wired electrocardiogram monitor in our operating room.

METHODS: Bluetooth electrocardiogram and standard electrocardiogram traces captured from 10 patients undergoing surgical procedures that required use of an electrosurgical unit were compared by analysis of the durations of the P wave, QRS complex, and T wave and the position of the ST segment from the isoelectric line. The impact of the electrosurgical units on the Bluetooth electrocardiogram and S-electrocardiogram recordings was also assessed.

RESULTS: There were no clinically relevant differences in P wave, QRS complex, or T-wave durations (0.006, 0.004, and 0.017 seconds, respectively) between Bluetooth electrocardiogram and standard electrocardiogram or in the position of the ST segment from the isoelectric line (0.02 mV). Mean differences were near zero, and Bland-Altman limits of agreement for individual differences were narrow (-0.035 to 0.047, -0.03 to 0.038, and -0.112 to 0.078 seconds for P wave, QRS complex, and T-wave durations, respectively, and -0.13 to 0.17 mV for ST segment position). Electrosurgical units use electrically disrupted Bluetooth electrocardiogram and standard electrocardiogram signals, but there was no electromagnetic interference effect on the Bluetooth electrocardiogram signals.

CONCLUSIONS: Wireless electrocardiogram using Bluetooth can be reliably used in the operating room. The electrosurgical unit induces electric rather than electromagnetic artifacts, thus affecting wired and wireless electrocardiogram in a similar fashion.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app