Quarter-millimeter spectral coronary stent imaging with photon-counting CT: Initial experience.
Journal of Cardiovascular Computed Tomography 2018 November
PURPOSE: To evaluate the performance and clinical feasibility of 0.25 mm resolution mode of a dual-energy photon-counting detector (PCD) computed tomography (CT) system for coronary stent imaging and to compare the results to state-of-the-art dual-energy energy-integrating detector (EID) CT.
MATERIALS AND METHODS: Coronary stents with different diameters (2.0-4.0 mm) were examined inside a coronary artery phantom consisting of plastic tubes filled with iodine-based and gadolinium-based contrast material diluted to approximate clinical concentrations (n = 18). EID images were acquired using 2nd and 3rd generation dual-source CT systems (SOMATOM Flash and SOMATOM Force, Siemens Healthcare) at 0.60 mm (defined as standard-resolution (SR)) isotropic voxel size. Radiation-dose matched PCD images were acquired using a human prototype PCD system (Siemens Healthcare) at 0.50 mm (SR) and 0.25 mm (HR) imaging modes. Images were reconstructed using optimized convolution kernels.
RESULTS: Dual-energy HR PCD images significantly better stent lumen visualization (median: 69.5%, IQR: 61.2-78.9%) over dual-energy EID, and standard-resolution PCD images (median: 53.2-57.4%, all P < 0.01). HR PCD acquisitions reconstructed at SR image voxel size showed 25.3% lower image noise compared to SR PCD acquisitions (P < 0.001). High-resolution iodine and gadolinium maps, as well as virtual monoenergetic images, were calculated from the PCD data and enabled estimation of contrast agent concentration in the lumen without interference from the coronary stent.
CONCLUSION: HR spectral PCD imaging significantly improves coronary stent lumen visibility over dual-energy EID. When the PCD-HR data was reconstructed into standard voxel sizes (0.5 mm isotropic) the image noise decreased by 25% compared to SR acquisition of PCD. Both dual-energy systems were consistent in estimating contrast agent concentrations.
MATERIALS AND METHODS: Coronary stents with different diameters (2.0-4.0 mm) were examined inside a coronary artery phantom consisting of plastic tubes filled with iodine-based and gadolinium-based contrast material diluted to approximate clinical concentrations (n = 18). EID images were acquired using 2nd and 3rd generation dual-source CT systems (SOMATOM Flash and SOMATOM Force, Siemens Healthcare) at 0.60 mm (defined as standard-resolution (SR)) isotropic voxel size. Radiation-dose matched PCD images were acquired using a human prototype PCD system (Siemens Healthcare) at 0.50 mm (SR) and 0.25 mm (HR) imaging modes. Images were reconstructed using optimized convolution kernels.
RESULTS: Dual-energy HR PCD images significantly better stent lumen visualization (median: 69.5%, IQR: 61.2-78.9%) over dual-energy EID, and standard-resolution PCD images (median: 53.2-57.4%, all P < 0.01). HR PCD acquisitions reconstructed at SR image voxel size showed 25.3% lower image noise compared to SR PCD acquisitions (P < 0.001). High-resolution iodine and gadolinium maps, as well as virtual monoenergetic images, were calculated from the PCD data and enabled estimation of contrast agent concentration in the lumen without interference from the coronary stent.
CONCLUSION: HR spectral PCD imaging significantly improves coronary stent lumen visibility over dual-energy EID. When the PCD-HR data was reconstructed into standard voxel sizes (0.5 mm isotropic) the image noise decreased by 25% compared to SR acquisition of PCD. Both dual-energy systems were consistent in estimating contrast agent concentrations.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Essential thrombocythaemia: A contemporary approach with new drugs on the horizon.British Journal of Haematology 2024 April 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app