How Much Surface Registration Accuracy is Required Using Ct-based Navigation System in Adolescent Idiopathic Scoliosis Surgery?

Clinical Spine Surgery 2018 November 31
STUDY DESIGN: Retrospective, single-center, single-surgeon study.

OBJECTIVE: This study investigated screw perforation rate according to surface registration accuracy in pedicle screw fixation using a CT-based navigation system for adolescent idiopathic scoliosis (AIS).

SUMMARY OF BACKGROUND DATA: Posterior spinal fusion for scoliosis correction using pedicle screws is widely adopted but carries a risk of serious neurovascular or visceral structure events. To attempt to avoid these complications, we have been using a computed tomography (CT)-based navigation system during screw insertion, although insufficient surface registration accuracy may be a reason for some screw perforations.

METHODS: We evaluated 116 AIS patients (9 male and 107 females; mean±standard deviation age: 15.2±2.3 y) who had received pedicle screw fixation guided by a CT-based navigation system between August 2007 and December 2017. Screw perforation rate was evaluated in relation to surface registration accuracy and cut-off values were determined by receiver operating characteristic (ROC) curves.

RESULTS: A total of 1553 pedicle screws were inserted into T2-L4 vertebrae using CT-based navigation system. The respective overall perforation rates for grade 2 or 3 and grade 3 perforations by surface registration accuracy were 0.2 mm: 0% and 0%; 0.3 mm: 3.5% and 0%; 0.4 mm: 3.5% and 0.8%; 0.5 mm: 8.0% and 3.5%; 0.6 mm: 6.7% and 2.7%; 0.7 mm: 8.9% and 3.8%; 0.8 mm: 9.3% and 6.7%; and 0.9 mm: 9.3% and 4.7%. ROC curve analysis revealed a cut-off surface registration accuracy of 0.5 mm. The major screw perforation rate for a surface registration accuracy of ≥0.5 mm was significantly higher than that for <0.5 mm (P<0.01).

CONCLUSIONS: Analysis of screw perforation rate according to surface registration accuracy demonstrated a clinical cut-off of 0.5 mm. Thus, surface registration accuracy should optimally be <0.5 mm for scoliosis surgery with CT-based navigation system.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app