Impact of cannula design on packed red blood cell transfusions: technical advancement to improve outcomes in extracorporeal membrane oxygenation.

Background: Technological improvement has contributed to making veno-venous extracorporeal membrane oxygenation (VV-ECMO) safer and easier, spreading its use in acute respiratory failure (ARF).

Methods: This is a retrospective observational study carried out in the ECMO center at IRCCS-ISMETT, a medical center focused on end-stage organ failure treatment in Italy. We investigated the effect of different cannula designs on the amount of blood product transfused. Eighty-nine consecutive patients affected with ARF on VV-ECMO from 2008 to 2016 were compared according to type of cannulation: older percutaneous cannula (Standard group, 52 patients) and HLS© BIOLINE-coated, but with shorter drainage cannula (BIOLINE group, 37 patients).

Results: The two study groups were comparable in terms of baseline characteristics [age, body mass index (BMI), Simplified Acute Physiology Score (SAPS-II), Sequential Organ Failure Assessment (SOFA), Predicting Death For Severe ARDS on VV-ECMO (PRESERVE) score] and ECMO management [median hematocrit (Htc), platelet nadir, antithrombin III (AT III), heparin, activated partial thromboplastin time (APTT)]. In the BIOLINE group, a lower amount of packed red blood cells (pRBC) was transfused considering both total number [4 units, interquartile range (IQR) 1-9 vs. 12 units, IQR 5.5-21; P<0.01] and mL of pRBC/day of ECMO support (91, IQR 21-158 vs. 193.5, IQR 140.5-254; P<0.01). In the BIOLINE group, a trend in reduction of ECMO days (P=0.05) and length of intensive care unit (ICU) stay was found (P=0.06), but no differences in rates of ECMO weaning and ICU discharge were evidenced. The BIOLINE group constituted a saving of €1,295.20 per patient/treatment, counting the costs for cannulation and pRBC administration.

Conclusions: More biocompatible and shorter drainage cannula may represent one of the contributing factors to a reduction in transfusions and costs of VV-ECMO in the current ongoing technological improvement in ECMO.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app