Non-Stationary Model of Oblique X-Ray Incidence in Amorphous Selenium Detectors: II. Transfer Functions.

Medical Physics 2018 November 29
PURPOSE: One limitation of experimental techniques for quantifying resolution and noise in detectors is that the measurement is made in a region-of-interest (ROI). With theoretical modeling, these properties can be measured at a point, allowing for quantification of spatial anisotropy. This paper calculates non-stationary transfer functions for amorphous selenium (a-Se) detectors in breast imaging. We use this model to demonstrate the performance advantage of a "next-generation" tomosynthesis (NGT) system, which is capable of x-ray source motion with more degrees of freedom than a clinical tomosynthesis system.

METHODS: Using Swank's formulation, the optical transfer function (OTF) and pre-sampled noise power spectra (NPS) are determined based on the point spread function derived in Part 1. The modulation transfer function (MTF) is found from the normalized modulus of the OTF. To take into account the presence of digitization, the pre-sampled NPS is convolved with a 2D comb function, for which the period along each direction is the reciprocal of the detector element size. The detective quantum efficiency (DQE) is then determined from combined knowledge of the OTF and NPS.

RESULTS: First, the model is used to demonstrate the loss of image quality due to oblique x-ray incidence. The MTF is calculated along various polar angles, corresponding to different orientations of the input frequency. The MTF is independent of the incidence angle if the polar angle is perpendicular to the ray incidence direction. However, along other polar angles, oblique incidence results in MTF degradation at high frequencies. The MTF degradation is most substantial along the ray incidence direction. Unlike the MTF, the normalized NPS (NNPS) is independent of the incidence angle. To measure the relative signal-to-noise, the DQE is also calculated. Oblique incidence yields high-frequency DQE degradation, which is more pronounced than the MTF degradation. This arises because the DQE is proportionate with the square of the MTF. Ultimately, this model is used to evaluate how image quality varies over the detector area. For various projection images, we calculate the variation in the incidence angle over this area. With the NGT system, the source can be positioned in such a way that this variation is minimized, and hence the DQE exhibits less anisotropy. To achieve this improvement in image quality, the source needs to have a component of motion in the posteroanterior (PA) direction, which is perpendicular to the conventional direction of source motion in tomosynthesis.

CONCLUSIONS: In a-Se detectors, the DQE at high frequencies is degraded due to oblique incidence. The DQE degradation is more pronounced than the MTF degradation. This model is used to quantify the spatial variation in DQE over the detector area. The use of PA source motion is a strategy for minimizing this variation and thus improving the image quality. This article is protected by copyright. All rights reserved.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app