AIM: To assess blooming artifacts caused by root canal sealers in CBCT images compared with those that appeared in micro-CT scan images used as references.
MATERIALS AND METHOD: Thirty freshly extracted human mandibular central incisors were used. Root canals were prepared with nickel titanium files with an ISO size 40/0.06 taper and filled with a single cone (40/0.06 taper) and three different sealers. The samples were divided into the following three groups with 10 roots each: (I) AH Plus sealer; (II) Sure Seal Root; and (III) Total BC sealer. Teeth were scanned with the same voxel sizes (0.2 mm) in different CBCT devices and the micro-CT images were acquired as reference images.
RESULTS: Significantly different results in terms of blooming artifacts were detected between CBCT and micro-CT images, as well as among the CBCTs images. The canals filled with AH Plus sealer showed more blooming artifacts than those filled with bioceramic sealers (p < 0.05). Additionally, the worst blooming artifact was observed when the images were acquired with lower kilovoltage peak.
CONCLUSION: The appearance of blooming artifacts is dependent on sealer and CBCT, and their effects are significantly worse than they are in micro-CT images. The differential effect of different sealers and distinct CBCT protocols should be further investigated to enable the use of bioceramic sealers without a significant impact on post-treatment imaging.
CLINICAL RELEVANCE: Root canal sealers showed a different extent of blooming artifact in CBCT images. Hence, researchers and clinicians should be aware of these artifacts before conducting endodontic evaluations using CBCT images.
MATERIALS AND METHOD: Thirty freshly extracted human mandibular central incisors were used. Root canals were prepared with nickel titanium files with an ISO size 40/0.06 taper and filled with a single cone (40/0.06 taper) and three different sealers. The samples were divided into the following three groups with 10 roots each: (I) AH Plus sealer; (II) Sure Seal Root; and (III) Total BC sealer. Teeth were scanned with the same voxel sizes (0.2 mm) in different CBCT devices and the micro-CT images were acquired as reference images.
RESULTS: Significantly different results in terms of blooming artifacts were detected between CBCT and micro-CT images, as well as among the CBCTs images. The canals filled with AH Plus sealer showed more blooming artifacts than those filled with bioceramic sealers (p < 0.05). Additionally, the worst blooming artifact was observed when the images were acquired with lower kilovoltage peak.
CONCLUSION: The appearance of blooming artifacts is dependent on sealer and CBCT, and their effects are significantly worse than they are in micro-CT images. The differential effect of different sealers and distinct CBCT protocols should be further investigated to enable the use of bioceramic sealers without a significant impact on post-treatment imaging.
CLINICAL RELEVANCE: Root canal sealers showed a different extent of blooming artifact in CBCT images. Hence, researchers and clinicians should be aware of these artifacts before conducting endodontic evaluations using CBCT images.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app