Three-Dimensional Printing Facilitates Creation of a Biliary Endoscopy Phantom for Interventional Radiology-Operated Endoscopy Training.

PURPOSE: To create a three-dimensional endoscopic model of the biliary tract from magnetic-resonance cholangiopancreatography imaging and to evaluate its effectiveness as a tool for training in endoscopic biliary interventions.

MATERIALS AND METHODS: A magnetic-resonance cholangiopancreatography study was performed on a patient with biliary obstruction secondary to a distal bile duct cholangiocarcinoma. Using Vitrea, a three-dimensional volume-rendered image was created, and exported as a standard tessellated language file. The standard tessellated language model was then edited with MeshMixer. Three cylindrical entry ports were created. The ports were aligned and overlapped with the dominant ducts in three separate areas of the model and fused to the model. A 0.2 cm shell was created around the model and the model was hollowed. The ends of the ports were cut off, allowing access to the hollowed-out model. The model was printed at 125% scale to allow easy access with a 9.5-French (≤3.23 mm) endoscope. The model was printed using a Stratasys Dimension Elite Plus printer. After printing, the model was post-processed to remove support materials. A 10-question survey was administered to all trainees before and after use of the printed phantom to practice endoscopy skills.

RESULTS: 11 trainees participated in the three-dimensional endoscopy simulation with most of the trainees (73%) having no prior formal endoscopy training. Using a 10-point Likert scale, the mean comfort-level of the trainees to use endoscopy alone for cholecystostomy, percutaneous biliary drainage, percutaneous nephrostomy, and percutaneous gastrostomy increased by 38.9%, 32.8%, 32.8%, and 34.3%, respectively, following the training experience.

CONCLUSION: The use of a three-dimensionally printed endoscopic model as a simulation tool has the potential to improve trainee comfort using endoscopy during interventional radiology procedures.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app