Six-Month Performance of a 3-Dimensional Annuloplasty Ring for Repair of Functional Tricuspid Regurgitation.
Thoracic and Cardiovascular Surgeon 2018 November 17
BACKGROUND: Functional tricuspid regurgitation (FTR) secondary to left-sided heart disease may lead to poor quality of life and reduced long-term survival. This study evaluated clinical and functional outcomes of patients undergoing tricuspid valve (TV) repair using a rigid three-dimensional ring (Contour 3D, Medtronic) concomitant with another procedure.
METHODS: From September 2011 to July 2015, 112 patients (mean age 70.9 ± 9.0 years) were enrolled at 10 centers in Europe, Israel, and the United States. Inclusion criteria were FTR ≥ moderate and/or tricuspid annular diameter (TAD) ≥ 40 mm. Echocardiography was planned before surgery and at discharge with echocardiographic and clinical follow-ups performed 6 months postoperatively.
RESULTS: Three fourths (74.4%) of patients had higher than moderate TR. Mean TAD was 41.0 ± 7.3 mm; 61.7% of patients were in the New York Heart Association (NYHA) class III/IV. The most common concomitant procedure was mitral valve repair (57 patients, 53.3%). The 30-day mortality rate was 0.9% ( n = 1). The mean EuroSCORE II was 8.9 ± 8.4% (median: 5.9%; interquartile range: 3.5-11.5%). The observed to expected ratio (O/E) based on the median was 0.1. Six deaths occurred during follow-up (three cardiac related). Mean implanted ring size was 30.3 ± 2.7. At 6 months, 94.4% of patients showed ≤ mild TR, and 92.0% were in NYHA class I/II ( p < 0.001 vs baseline for both). Mean pressure gradient across the TV was 2.0 ± 1.1 mm Hg; leaflet coaptation length was 7.5 ± 3.3 mm.
CONCLUSION: The Contour 3D annuloplasty ring used for treatment of FTR substantially reduced TR for up to 6 postoperative months with low mean pressure gradients across the TV and significant improvement in NYHA class.
REGISTRATION: www.ClinicalTrials.gov, NCT01532921.
METHODS: From September 2011 to July 2015, 112 patients (mean age 70.9 ± 9.0 years) were enrolled at 10 centers in Europe, Israel, and the United States. Inclusion criteria were FTR ≥ moderate and/or tricuspid annular diameter (TAD) ≥ 40 mm. Echocardiography was planned before surgery and at discharge with echocardiographic and clinical follow-ups performed 6 months postoperatively.
RESULTS: Three fourths (74.4%) of patients had higher than moderate TR. Mean TAD was 41.0 ± 7.3 mm; 61.7% of patients were in the New York Heart Association (NYHA) class III/IV. The most common concomitant procedure was mitral valve repair (57 patients, 53.3%). The 30-day mortality rate was 0.9% ( n = 1). The mean EuroSCORE II was 8.9 ± 8.4% (median: 5.9%; interquartile range: 3.5-11.5%). The observed to expected ratio (O/E) based on the median was 0.1. Six deaths occurred during follow-up (three cardiac related). Mean implanted ring size was 30.3 ± 2.7. At 6 months, 94.4% of patients showed ≤ mild TR, and 92.0% were in NYHA class I/II ( p < 0.001 vs baseline for both). Mean pressure gradient across the TV was 2.0 ± 1.1 mm Hg; leaflet coaptation length was 7.5 ± 3.3 mm.
CONCLUSION: The Contour 3D annuloplasty ring used for treatment of FTR substantially reduced TR for up to 6 postoperative months with low mean pressure gradients across the TV and significant improvement in NYHA class.
REGISTRATION: www.ClinicalTrials.gov, NCT01532921.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app