We have located links that may give you full text access.
A practical approach to estimating optic disc dose and macula dose without treatment planning in ocular brachytherapy using 125 I COMS plaques.
Radiation Oncology 2018 November 14
BACKGROUND: It has been reported that proximity of the tumor to the optic disc and macula, and radiation dose to the critical structures are substantial risk factors for vision loss following plaque brachytherapy. However, there is little dosimetry data published on this. In this study, therefore, the relationship between distance from tumor margin and radiation dose to the optic disc and macula in ocular brachytherapy using 125 I Collaborative Ocular Melanoma Study (COMS) plaques was comprehensively investigated. From the information, this study aimed to allow for estimation of optic disc dose and macula dose without treatment planning.
METHODS: An in-house brachytherapy dose calculation program utilizing the American Association of Physicists in Medicine Task Group-43 U1 formalism with a line source approximation in a homogenous water phantom was developed and validated against three commercial treatment planning systems (TPS). Then optic disc dose and macula dose were calculated as a function of distance from tumor margin for various tumor basal dimensions for seven COMS plaques (from 10 mm to 22 mm in 2 mm increments) loaded with commercially available 125 I seeds models (IAI-125A, 2301 and I25.S16). A prescribed dose of 85 Gy for an irradiation time of 168 h was normalized to a central-axis depth of 5 mm. Dose conversion factors for each seed model were obtained by taking ratios of total reference air kerma per seed at various prescription depths (from 1 mm to 10 mm in 1 mm intervals) to that at 5 mm.
RESULTS: The in-house program demonstrated relatively similar accuracy to commercial TPS. Optic disc dose and macula dose decreased as distance from tumor margin and tumor basal dimension increased. Dose conversion factors increased with increasing prescription depth. There existed dose variations (<8%) among three 125 I seed models. Optic disc dose and macula dose for each COMS plaque and for each seed model are presented in a figure format. Dose conversion factors for each seed model are presented in a tabular format.
CONCLUSIONS: The data provided in this study would enable clinicians in any clinic using 125 I COMS plaques to estimate optic disc dose and macula dose without dose calculations.
METHODS: An in-house brachytherapy dose calculation program utilizing the American Association of Physicists in Medicine Task Group-43 U1 formalism with a line source approximation in a homogenous water phantom was developed and validated against three commercial treatment planning systems (TPS). Then optic disc dose and macula dose were calculated as a function of distance from tumor margin for various tumor basal dimensions for seven COMS plaques (from 10 mm to 22 mm in 2 mm increments) loaded with commercially available 125 I seeds models (IAI-125A, 2301 and I25.S16). A prescribed dose of 85 Gy for an irradiation time of 168 h was normalized to a central-axis depth of 5 mm. Dose conversion factors for each seed model were obtained by taking ratios of total reference air kerma per seed at various prescription depths (from 1 mm to 10 mm in 1 mm intervals) to that at 5 mm.
RESULTS: The in-house program demonstrated relatively similar accuracy to commercial TPS. Optic disc dose and macula dose decreased as distance from tumor margin and tumor basal dimension increased. Dose conversion factors increased with increasing prescription depth. There existed dose variations (<8%) among three 125 I seed models. Optic disc dose and macula dose for each COMS plaque and for each seed model are presented in a figure format. Dose conversion factors for each seed model are presented in a tabular format.
CONCLUSIONS: The data provided in this study would enable clinicians in any clinic using 125 I COMS plaques to estimate optic disc dose and macula dose without dose calculations.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app