How to perform the dusting technique for calcium oxalate stone phantoms during Ho:YAG laser lithotripsy.
BMC Urology 2018 November 14
BACKGROUND: To determine the most efficacious setting of Holmium:yttrium-aluminum-garnet (Ho:YAG) laser with a maximum power output of 120 W with in vitro phantom-stone dusting technique.
METHODS: A laser was used to treat two 4 × 3 × 3 mm3 sized phantom stones in 5 mL syringes with 1 mm-sized holes at the bottom. According to the pulse width (short 500, middle 750, long pulse 1000 μsec), maximal pulse repetition rates from 50 to 80 Hz were tested with pulse energy of 0.2, 0.4, 0.5, and 0.8 J. Six times of the mean dusting times were measured at each setting. Dusting was performed at continuous firing of the laser until the stones become dusts < 1 mm.
RESULTS: The mean Hounsfield unit of phantom stones was 1309.0 ± 60.8. The laser with long pulse generally showed shorter dusting times than short or middle pulse width. With increasing the pulse energy to 0.5 J, the dusting time decreased. However, the pulse energy of 0.8 J showed longer dusting times than those of 0.5 J. On the post-hoc analysis, the pulse energy of 0.5 J, long pulse width, and the repetition rates of 70 Hz demonstrated significantly shorter dusting times than other settings.
CONCLUSIONS: The results suggest that long pulse width with 0.5 J and 70 Hz would be the most efficacious setting for dusting techniques of plaster stone phantoms simulating calcium oxalate stones using the 120 W Ho:YAG laser.
METHODS: A laser was used to treat two 4 × 3 × 3 mm3 sized phantom stones in 5 mL syringes with 1 mm-sized holes at the bottom. According to the pulse width (short 500, middle 750, long pulse 1000 μsec), maximal pulse repetition rates from 50 to 80 Hz were tested with pulse energy of 0.2, 0.4, 0.5, and 0.8 J. Six times of the mean dusting times were measured at each setting. Dusting was performed at continuous firing of the laser until the stones become dusts < 1 mm.
RESULTS: The mean Hounsfield unit of phantom stones was 1309.0 ± 60.8. The laser with long pulse generally showed shorter dusting times than short or middle pulse width. With increasing the pulse energy to 0.5 J, the dusting time decreased. However, the pulse energy of 0.8 J showed longer dusting times than those of 0.5 J. On the post-hoc analysis, the pulse energy of 0.5 J, long pulse width, and the repetition rates of 70 Hz demonstrated significantly shorter dusting times than other settings.
CONCLUSIONS: The results suggest that long pulse width with 0.5 J and 70 Hz would be the most efficacious setting for dusting techniques of plaster stone phantoms simulating calcium oxalate stones using the 120 W Ho:YAG laser.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app