Analytical and Clinical Performance Evaluation of the Elecsys HIV combi PT Assay on the cobas e 602 Analyzer for the Diagnosis of Human Immunodeficiency Virus.
American Journal of Clinical Pathology 2018 November 14
Objectives: We evaluated the performance of the Elecsys HIV combi PT assay on the cobas e 602 analyzer for diagnosing human immunodeficiency virus (HIV; part of the US Food and Drug Administration [FDA] submission).
Methods: The HIV combi PT and reference (ARCHITECT HIV Ag/Ab Combo) assays were assessed at four independent clinical laboratories/one reference laboratory (United States; July 2014 to November 2015). Clinical performance was evaluated using four reagent lots. Analytical performance was evaluated per Clinical and Laboratory Standards Institute EP05-A3 guidelines. Serum/plasma samples from 18 clinical sites/vendors (United States and outside the United States) were tested.
Results: Sensitivity (95% confidence interval [CI]) in HIV-1 antibody-positive individuals (United States and outside the United States; n = 1,460) was 100.00% (99.75%-100.00%). Specificity was 99.94% (95% CI, 99.85%-99.98%) in low-risk individuals (United States; n = 6,843), 98.19% (95% CI, 96.93%-99.04%) in high-risk individuals (United States and outside the United States; n = 758), and 97.43% (95% CI, 95.32%-98.76%) in pregnant women (United States and outside the United States; n = 440). Analytical performance was acceptable.
Conclusions: We demonstrate the robustness of the FDA-approved Elecsys HIV combi PT assay on the cobas e 602 analyzer for HIV testing in the United States.
Methods: The HIV combi PT and reference (ARCHITECT HIV Ag/Ab Combo) assays were assessed at four independent clinical laboratories/one reference laboratory (United States; July 2014 to November 2015). Clinical performance was evaluated using four reagent lots. Analytical performance was evaluated per Clinical and Laboratory Standards Institute EP05-A3 guidelines. Serum/plasma samples from 18 clinical sites/vendors (United States and outside the United States) were tested.
Results: Sensitivity (95% confidence interval [CI]) in HIV-1 antibody-positive individuals (United States and outside the United States; n = 1,460) was 100.00% (99.75%-100.00%). Specificity was 99.94% (95% CI, 99.85%-99.98%) in low-risk individuals (United States; n = 6,843), 98.19% (95% CI, 96.93%-99.04%) in high-risk individuals (United States and outside the United States; n = 758), and 97.43% (95% CI, 95.32%-98.76%) in pregnant women (United States and outside the United States; n = 440). Analytical performance was acceptable.
Conclusions: We demonstrate the robustness of the FDA-approved Elecsys HIV combi PT assay on the cobas e 602 analyzer for HIV testing in the United States.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app