Reproducibility of the durometer and myoton devices for skin stiffness measurement in healthy subjects.
Skin Research and Technology 2018 November 11
BACKGROUND: Clinical assessment of skin stiffness is unreliable in many applications. The durometer, an industrial device to measure hardness, has previously been applied in scleroderma. The Myoton is a noninvasive handheld device for assessing soft tissue biomechanical parameters.
MATERIALS AND METHODS: We evaluated the reproducibility of both devices in six healthy subjects in the volar forearm, dorsal forearm, upper arm, shin, and calf bilaterally. The intraclass correlation coefficient (ICC) was used as a measure of reproducibility among three observers.
RESULTS: The interobserver intraclass correlation coefficient (ICC) of overall stiffness for the Myoton was 0.74 [95% confidence interval (CI) 0.45-1.00] and 0.71 [0.39-1.00] for the durometer. Coefficient of variation (CV) for the Myoton was 6.4% [range 1.3-12.1] and 7.6% [range 4.4-13.8] for the durometer. Myoton and durometer values had a Pearson correlation of 0.69. The intraobserver Myoton ICC was 0.89 [0.74-1.00] and CV 3.1% [range 1.6-5.0]. The 95% confidence minimal detectable change by the Myoton for a single observer is 32.4 N/m, which is 7.6% of the average subject's overall stiffness.
CONCLUSION: The Myoton demonstrated high reproducibility, particularly in the overall stiffness parameter, and merits further investigation to assess disease progression and treatment efficacy.
MATERIALS AND METHODS: We evaluated the reproducibility of both devices in six healthy subjects in the volar forearm, dorsal forearm, upper arm, shin, and calf bilaterally. The intraclass correlation coefficient (ICC) was used as a measure of reproducibility among three observers.
RESULTS: The interobserver intraclass correlation coefficient (ICC) of overall stiffness for the Myoton was 0.74 [95% confidence interval (CI) 0.45-1.00] and 0.71 [0.39-1.00] for the durometer. Coefficient of variation (CV) for the Myoton was 6.4% [range 1.3-12.1] and 7.6% [range 4.4-13.8] for the durometer. Myoton and durometer values had a Pearson correlation of 0.69. The intraobserver Myoton ICC was 0.89 [0.74-1.00] and CV 3.1% [range 1.6-5.0]. The 95% confidence minimal detectable change by the Myoton for a single observer is 32.4 N/m, which is 7.6% of the average subject's overall stiffness.
CONCLUSION: The Myoton demonstrated high reproducibility, particularly in the overall stiffness parameter, and merits further investigation to assess disease progression and treatment efficacy.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Essential thrombocythaemia: A contemporary approach with new drugs on the horizon.British Journal of Haematology 2024 April 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app