Leadless Cardiac Pacemaker (LCP) without Diagnostic Relevant Artifacts in DualSource and DualEnergy-CT Examinations in First- to Third-Generation DSCT Scanner.

Academic Radiology 2018 November 7
RATIONALE AND OBJECTIVES: To identify the influence and artifact burden in cardiac CT imaging of a leadless cardiac pacemaker (LCP) performed with all three generations of DualSource CT (DSCT) Scanners.

MATERIALS AND METHODS: The LCP was examined in DSCT scanners of the first to third generation using DualEnergy (DECT) and DSCT as well as alterations of the current-time product. For DECT examinations, virtual monoenergetic images were computed manually on a dedicated workstation. Virtual voltage was manually selected by subjective assessment of the lowest artifact burden. Systematic variations of the pacemaker angle to the gantry were assessed, too. The angle was successively increased by 10°, ranging from 0° to 90°. Artifact burden was quantified on a five-point Likert scale (1- no artifacts, 2- few artifacts, 3- moderate artifacts, 4- many artifacts, and 5- massive artifacts). Likert values of 1-3 were considered diagnostic and assessed by two board-certified radiologists in consensus.

RESULTS: In total, 200 examinations were analyzed, a mean Likert value of 1.93 ± 0.61 was found overall. None of the images were assessed Likert value >3. The positioning evaluation showed a clear and significant reduction of artifact burden toward lower angles, (0°: 1.4 ± 0.5 vs. 90° 2.55 ± 0.51). At scanner level, second-generation DSCT performed significantly better (1.68 ± 0.47) than both other scanners. Comparison of technique (DECT vs. DSCT) revealed a significantly improved image quality in DSCT examinations.

CONCLUSION: LCP can be safely examined in DSCT scanner of the first to third generation with the evaluated protocols and techniques, which are currently in use. Artifact burden can be significantly reduced by aligning or approaching the LCP's longitudinal axis toward the scanner's z-axis.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app