Point-of-care measurements of blood ketones in newborns.
Archives of Disease in Childhood. Fetal and Neonatal Edition 2019 September
OBJECTIVE: Babies may use alternative cerebral fuels including ketones when blood glucose concentrations are low, but laboratory ketone measurements are slow and expensive. Point-of-care measurement of ketone concentrations, if sufficiently accurate, may provide useful information for clinical care.
PATIENTS AND DESIGN: Eligible babies were 35-42 weeks' gestation, ≤10 days old and admitted to the newborn intensive care unit. At the time of clinically indicated blood tests, additional samples were taken to measure beta-hydroxybutyrate using a point-of-care analyser and the laboratory method.
RESULTS: One-hundred and fifty babies had 142 paired samples. Overall point-of-care accuracy was excellent (mean difference 0.00 mmol/L) and precision was moderate (SD 0.18 mmol/L). A point-of-care measurement ≥0.4 mmol/L was highly predictive of a laboratory measurement ≥0.4 mmol/L (area under the curve 0.98).
CONCLUSION: Point-of-care measurement of blood beta-hydroxybutyrate concentrations is sufficiently accurate in newborns to be potentially useful in clinical care.
CLINICAL TRIAL REGISTRATION NUMBER: Registered with the Australian and New Zealand Clinical Trials Registry ACTRN: 12616000784415. The study was registered before recruitment commenced.
PATIENTS AND DESIGN: Eligible babies were 35-42 weeks' gestation, ≤10 days old and admitted to the newborn intensive care unit. At the time of clinically indicated blood tests, additional samples were taken to measure beta-hydroxybutyrate using a point-of-care analyser and the laboratory method.
RESULTS: One-hundred and fifty babies had 142 paired samples. Overall point-of-care accuracy was excellent (mean difference 0.00 mmol/L) and precision was moderate (SD 0.18 mmol/L). A point-of-care measurement ≥0.4 mmol/L was highly predictive of a laboratory measurement ≥0.4 mmol/L (area under the curve 0.98).
CONCLUSION: Point-of-care measurement of blood beta-hydroxybutyrate concentrations is sufficiently accurate in newborns to be potentially useful in clinical care.
CLINICAL TRIAL REGISTRATION NUMBER: Registered with the Australian and New Zealand Clinical Trials Registry ACTRN: 12616000784415. The study was registered before recruitment commenced.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app