Stability of Values for the Activities of Critical Enzymes Assayed in Serum Frozen for Prolonged Time Periods.

OBJECTIVES: Our medical center laboratory receives frozen clinical chemistry samples from outlying hospitals for which assays for critical enzyme activities are requested. Our objective is to determine the effects of freezing (-20°C) on these enzyme activities in samples over a one month period.

METHODS: Enzyme activities for ALP, AST, ALT, CK and LD for 30 patient's sera were stored at-20°C and were assayed on a Beckman-Coulter AU5800 analyzer at 0, 15 and 30 days after collection. Statistical tests were performed to determine if the values were statistically the same or different.

RESULTS: F-tests for all five enzyme levels showed no statistically significant differences ( p >0.05); linear regression analysis showed high correlation of results (r>0.99 for all correlations) with some bias for ALT.

CONCLUSIONS: We conclude that the activities of these enzymes are stable, except possibly ALT, when stored frozen at -20°C over the 15- and 30-day storage periods.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app