Ultrashort echo time imaging for quantification of hepatic iron overload: Comparison of acquisition and fitting methods via simulations, phantoms, and in vivo data.
Journal of Magnetic Resonance Imaging : JMRI 2018 October 26
BACKGROUND: Current R2*-MRI techniques for measuring hepatic iron content (HIC) use various acquisition types and fitting models.
PURPOSE: To evaluate the accuracy and precision of R2*-HIC acquisition and fitting methods.
STUDY TYPE: Signal simulations, phantom study, and prospective in vivo cohort.
POPULATION: In all, 132 patients (58/74 male/female, mean age 17.7 years).
FIELD STRENGTH/SEQUENCE: 2D-multiecho gradient-echo (GRE) and ultrashort echo time (UTE) acquisitions at 1.5T.
ASSESSMENT: Synthetic MR signals were created to mimic published GRE and UTE methods, using different R2* values (25-2000 s-1 ) and signal-to-noise ratios (SNR). Phantoms with varying iron concentrations were scanned at 1.5T. In vivo data were analyzed from 132 patients acquired at 1.5T. R2* was estimated by fitting using three signal models. Accuracy and precision of R2* measurements for UTE acquisition parameters (SNR, echo spacing [ΔTE], maximum echo time [TEmax ]) and fitting methods were compared for simulated, phantom, and in vivo datasets.
STATISTICAL TESTS: R2* accuracy was determined from the relative error and by linear regression analysis. Precision was evaluated using coefficient of variation (CoV) analysis.
RESULTS: In simulations, all models had high R2* accuracy (error <5%) and precision (CoV <10%) for all SNRs, shorter ΔTE (≤0.5 msec), and longer TEmax (≥10.1 msec); except the constant offset model overestimated R2* at the lowest SNR. In phantoms and in vivo, all models produced similar R2* values for different SNRs and shorter ΔTEs (slopes: 0.99-1.06, R2 > 0.99, P < 0.001). In all experiments, R2* results degraded for high R2* values with longer ΔTE (≥1 msec). In vivo, shorter and longer TEmax gave similar R2* results (slopes: 1.02-1.06, R2 > 0.99, P < 0.001) for the noise subtraction model for 25≤R2*≤2000 s-1 . However, both quadratic and constant offset models, using shorter TEmax (≤4.7 msec) overestimated R2* and yielded high CoVs up to ∼170% for low R2* (<250 s-1 ).
DATA CONCLUSION: UTE with TEmax ≥ 10.1 msec and ΔTE ≤ 0.5 msec yields accurate R2* estimates over the entire clinical HIC range. Monoexponential fitting with noise subtraction is the most robust signal model to changes in UTE parameters and achieves the highest R2* accuracy and precision.
LEVEL OF EVIDENCE: 2 Technical Efficacy: Stage 2.
PURPOSE: To evaluate the accuracy and precision of R2*-HIC acquisition and fitting methods.
STUDY TYPE: Signal simulations, phantom study, and prospective in vivo cohort.
POPULATION: In all, 132 patients (58/74 male/female, mean age 17.7 years).
FIELD STRENGTH/SEQUENCE: 2D-multiecho gradient-echo (GRE) and ultrashort echo time (UTE) acquisitions at 1.5T.
ASSESSMENT: Synthetic MR signals were created to mimic published GRE and UTE methods, using different R2* values (25-2000 s-1 ) and signal-to-noise ratios (SNR). Phantoms with varying iron concentrations were scanned at 1.5T. In vivo data were analyzed from 132 patients acquired at 1.5T. R2* was estimated by fitting using three signal models. Accuracy and precision of R2* measurements for UTE acquisition parameters (SNR, echo spacing [ΔTE], maximum echo time [TEmax ]) and fitting methods were compared for simulated, phantom, and in vivo datasets.
STATISTICAL TESTS: R2* accuracy was determined from the relative error and by linear regression analysis. Precision was evaluated using coefficient of variation (CoV) analysis.
RESULTS: In simulations, all models had high R2* accuracy (error <5%) and precision (CoV <10%) for all SNRs, shorter ΔTE (≤0.5 msec), and longer TEmax (≥10.1 msec); except the constant offset model overestimated R2* at the lowest SNR. In phantoms and in vivo, all models produced similar R2* values for different SNRs and shorter ΔTEs (slopes: 0.99-1.06, R2 > 0.99, P < 0.001). In all experiments, R2* results degraded for high R2* values with longer ΔTE (≥1 msec). In vivo, shorter and longer TEmax gave similar R2* results (slopes: 1.02-1.06, R2 > 0.99, P < 0.001) for the noise subtraction model for 25≤R2*≤2000 s-1 . However, both quadratic and constant offset models, using shorter TEmax (≤4.7 msec) overestimated R2* and yielded high CoVs up to ∼170% for low R2* (<250 s-1 ).
DATA CONCLUSION: UTE with TEmax ≥ 10.1 msec and ΔTE ≤ 0.5 msec yields accurate R2* estimates over the entire clinical HIC range. Monoexponential fitting with noise subtraction is the most robust signal model to changes in UTE parameters and achieves the highest R2* accuracy and precision.
LEVEL OF EVIDENCE: 2 Technical Efficacy: Stage 2.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app