Prescreening for Osteoporosis With Quantitative Ultrasound in Postmenopausal White Women.
OBJECTIVES: Calcaneal quantitative ultrasound (QUS) is a readily accessible and radiation-free alternative to dual-energy x-ray absorptiometry (DXA) for assessing bone mineral density (BMD). Results obtained from QUS measurement cannot directly be compared to DXA, since these techniques capture different bone-specific parameters. To identify individuals who are likely to have osteoporosis by DXA, device-specific thresholds have to be defined for QUS. This cross-sectional study evaluated the accuracy of QUS to identify postmenopausal women with osteoporosis, defined as a T score of -2.5 SDs or lower by DXA, and to calculate device-specific cutoff values for the QUS device investigated.
METHODS: We assessed BMD at the lumbar spine, bilateral femoral neck, and total hip sites with DXA and QUS parameters of the right and left calcanei in a cohort of 245 postmenopausal treatment-naïve women between 40 and 82 years. Correlation coefficients for BMD and QUS parameters were calculated. Receiver operating characteristic curves were generated, and areas under the curves (AUCs) were evaluated. Cutoff values for QUS were defined.
RESULTS: Calcaneal QUS had the ability to identify postmenopausal women with a T score of -2.5 or lower at the right hip (AUC, 0.887) and left femoral neck (AUC, 0.824). Cutoff values for the QUS T scores at the right (-1.455) and left (-1.480) calcanei were defined for screening purposes.
CONCLUSIONS: This study provides insights into the comparative performance of QUS with DXA. Considering the diagnostic accuracy of this modality in comparison to DXA, it can be recommended as a prescreening tool to reduce the number of DXA screenings.
METHODS: We assessed BMD at the lumbar spine, bilateral femoral neck, and total hip sites with DXA and QUS parameters of the right and left calcanei in a cohort of 245 postmenopausal treatment-naïve women between 40 and 82 years. Correlation coefficients for BMD and QUS parameters were calculated. Receiver operating characteristic curves were generated, and areas under the curves (AUCs) were evaluated. Cutoff values for QUS were defined.
RESULTS: Calcaneal QUS had the ability to identify postmenopausal women with a T score of -2.5 or lower at the right hip (AUC, 0.887) and left femoral neck (AUC, 0.824). Cutoff values for the QUS T scores at the right (-1.455) and left (-1.480) calcanei were defined for screening purposes.
CONCLUSIONS: This study provides insights into the comparative performance of QUS with DXA. Considering the diagnostic accuracy of this modality in comparison to DXA, it can be recommended as a prescreening tool to reduce the number of DXA screenings.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Essential thrombocythaemia: A contemporary approach with new drugs on the horizon.British Journal of Haematology 2024 April 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app