The effectiveness of additional lead-shielding drape and low pulse rate fluoroscopy in protecting staff from scatter radiation during cardiac resynchronization therapy (CRT).
Japanese Journal of Radiology 2018 October 16
PURPOSE: Cardiac resynchronization therapy (CRT) often requires a long fluoroscopic time and protection from scatter radiation. This study reports on scatter radiation levels during CRT, with and without additional shielding, and using standard or low pulse rate fluoroscopy.
MATERIALS AND METHODS: Additional lead-shielding drape (0.35-mm lead equivalent) was used on the left side of the table and pulsed fluoroscopy was performed at rates of 10 pulses/s (usual rate) and 7.5 pulses/s (low pulse rate). Fluoroscopy scatter radiation was measured for both pulse rates using an acrylic phantom with a radiation survey meter, both with and without the additional lead-shielding drape.
RESULTS: With the additional lead-shielding drape, the fluoroscopy scatter radiation was reduced by 74.3% at 10 pulses/s and 78.6% at 7.5 pulses/s. If the fluoroscopy was changed from 10 pulses/s to 7.5 pulses/s, the scattered radiation at the primary physician's position was reduced by 24.0%. The combined use of additional shielding drape and low pulse rate fluoroscopy reduced scatter radiation by over 80%.
CONCLUSION: Additional lead-shielding drape and low pulse rate fluoroscopy are effective in reducing the scattered radiation dose to physicians and nurses during CRT.
MATERIALS AND METHODS: Additional lead-shielding drape (0.35-mm lead equivalent) was used on the left side of the table and pulsed fluoroscopy was performed at rates of 10 pulses/s (usual rate) and 7.5 pulses/s (low pulse rate). Fluoroscopy scatter radiation was measured for both pulse rates using an acrylic phantom with a radiation survey meter, both with and without the additional lead-shielding drape.
RESULTS: With the additional lead-shielding drape, the fluoroscopy scatter radiation was reduced by 74.3% at 10 pulses/s and 78.6% at 7.5 pulses/s. If the fluoroscopy was changed from 10 pulses/s to 7.5 pulses/s, the scattered radiation at the primary physician's position was reduced by 24.0%. The combined use of additional shielding drape and low pulse rate fluoroscopy reduced scatter radiation by over 80%.
CONCLUSION: Additional lead-shielding drape and low pulse rate fluoroscopy are effective in reducing the scattered radiation dose to physicians and nurses during CRT.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Essential thrombocythaemia: A contemporary approach with new drugs on the horizon.British Journal of Haematology 2024 April 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app