Accuracy and Trending Ability of the Fourth-Generation FloTrac/EV1000 System in Patients With Severe Aortic Valve Stenosis Before and After Surgical Valve Replacement.
OBJECTIVE: Evaluate the accuracy and trending ability of the fourth-generation FloTrac/EV1000 (Edwards Lifesciences, Irvine, CA) system in patients with severe aortic valve stenosis by comparing FloTrac/EV1000-derived cardiac output (CCO-FT) with continuous thermodilution pulmonary artery catheter (CCO-PAC) measurements before and after surgical valve replacement.
DESIGN: Prospective clinical study.
SETTING: Anesthesia for cardiac surgery, operating room, single-center university hospital.
PARTICIPANTS: Twenty-five patients were included. After exclusion, 20 patients undergoing elective aortic valve replacement were analyzed.
INTERVENTIONS: After induction of general anesthesia, CCO-FT and CCO-PAC values were recorded every 30 seconds before and after aortic valve replacement with a bioprosthesis under cardiopulmonary bypass (CPB).
MEASUREMENTS AND MAIN RESULTS: Data were analyzed separately from skin incision to last suture and before and after CPB. Regression analyses, Bland-Altman analyses, and trending analyses (4-quadrant plot, polar plot) were performed. The percentage errors of the FloTrac/EV1000 were 69.7% and 59.3% before and after CPB, respectively. The concordance rates (CRs) and angular CRs of the FloTrac/EV1000 were 50.9% and 57.1%, and 48.7% and 61.9% before and after CPB, respectively.
CONCLUSION: This study revealed a low level of agreement and poor trending ability of the FloTrac/EV1000 system compared to continuous thermodilution pulmonary artery catheter in patients with severe aortic stenosis. Although there was a slight improvement after surgical valve replacement and CPB, the results were not within acceptable limits to replace CCO-PAC in this patient population.
DESIGN: Prospective clinical study.
SETTING: Anesthesia for cardiac surgery, operating room, single-center university hospital.
PARTICIPANTS: Twenty-five patients were included. After exclusion, 20 patients undergoing elective aortic valve replacement were analyzed.
INTERVENTIONS: After induction of general anesthesia, CCO-FT and CCO-PAC values were recorded every 30 seconds before and after aortic valve replacement with a bioprosthesis under cardiopulmonary bypass (CPB).
MEASUREMENTS AND MAIN RESULTS: Data were analyzed separately from skin incision to last suture and before and after CPB. Regression analyses, Bland-Altman analyses, and trending analyses (4-quadrant plot, polar plot) were performed. The percentage errors of the FloTrac/EV1000 were 69.7% and 59.3% before and after CPB, respectively. The concordance rates (CRs) and angular CRs of the FloTrac/EV1000 were 50.9% and 57.1%, and 48.7% and 61.9% before and after CPB, respectively.
CONCLUSION: This study revealed a low level of agreement and poor trending ability of the FloTrac/EV1000 system compared to continuous thermodilution pulmonary artery catheter in patients with severe aortic stenosis. Although there was a slight improvement after surgical valve replacement and CPB, the results were not within acceptable limits to replace CCO-PAC in this patient population.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app