Titanium particles generated during ultrasonic scaling of implants.
Journal of Periodontology 2018 October 13
BACKGROUND: There is growing concern that titanium particles may play a role in peri-implant breakdown. Ultrasonic scalers are routinely used in the debridement of peri-implant lesions. This in vitro study is designed to evaluate if titanium particles are produced when an ultrasonic scaler is used on an implant.
METHODS: New sandblasted, large grit, acid etched (SLA) coated implants were subjected to ultrasonic scaling with stainless steel, titanium, and PEEK plastic tips. The implants were placed in a holding device and the ultrasonic scaler was positioned on the SLA surface under 25 grams of pressure. The implants were subjected to 30 scaling motions. The ultrasonic coolant water was collected and the number of metallic particles were counted under a light microscope. The particles were confirmed to be titanium via elemental analysis. The implants were visually evaluated for damage to the SLA coating.
RESULTS: No metallic particles were detected in the water supplied to the ultrasonic scalers (passive control). Metallic particles were detected when implants were subjected to the ultrasonic coolant water only without the scaler tip touching the implant (active control). All implants that were scaled produced metallic particles and showed easily detectable damage to the SLA layer.
CONCLUSION: All ultrasonic scaling caused the production of titanium particles and caused damage to the SLA coating of the implant. Ultrasonic scalers should be used with great caution in the treatment of peri-implant conditions and care should be taken to not touch the SLA surface of the implant. This article is protected by copyright. All rights reserved.
METHODS: New sandblasted, large grit, acid etched (SLA) coated implants were subjected to ultrasonic scaling with stainless steel, titanium, and PEEK plastic tips. The implants were placed in a holding device and the ultrasonic scaler was positioned on the SLA surface under 25 grams of pressure. The implants were subjected to 30 scaling motions. The ultrasonic coolant water was collected and the number of metallic particles were counted under a light microscope. The particles were confirmed to be titanium via elemental analysis. The implants were visually evaluated for damage to the SLA coating.
RESULTS: No metallic particles were detected in the water supplied to the ultrasonic scalers (passive control). Metallic particles were detected when implants were subjected to the ultrasonic coolant water only without the scaler tip touching the implant (active control). All implants that were scaled produced metallic particles and showed easily detectable damage to the SLA layer.
CONCLUSION: All ultrasonic scaling caused the production of titanium particles and caused damage to the SLA coating of the implant. Ultrasonic scalers should be used with great caution in the treatment of peri-implant conditions and care should be taken to not touch the SLA surface of the implant. This article is protected by copyright. All rights reserved.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app