Reporting Insulin Pump Accuracy: Trumpet Curves According to IEC 60601-2-24 and Beyond.

Accuracy of insulin pump basal rate delivery, if tested according to the standard IEC 60601-2-24 for infusion pumps, shall be presented as a trumpet curve. This way of graphical presentation is common; however, it is often misunderstood and misinterpreted by people. It is often assumed that a trumpet curve shows the error rate as a function of time, thus implying an increasing accuracy in the course of time. On the contrary, the horizontal axis of a trumpet curve shows increasingly long observation windows. In addition, trumpet curves display only extreme values, that is, those windows with minimal and maximal deviation, which might not be representative for the total deviation. This commentary provides information regarding the calculations and the interpretation of trumpet curves and proposes alternative approaches.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app