Assessment of subjective refraction with a clinical adaptive optics visual simulator.
Journal of Cataract and Refractive Surgery 2018 October 9
PURPOSE: To clinically validate an adaptive optics visual simulator (VAO) that measures subjective refraction and visual acuity.
SETTING: Optics Laboratory, University of Murcia, Murcia, Spain.
DESIGN: Prospective case series.
METHODS: Using the adaptive optics visual simulator, 2 examiners measured the subjective refraction and visual acuity in healthy eyes of volunteers; 1 examiner also used a trial frame as a gold standard. The interexaminer reproducibility and agreement with the gold standard were estimated using the following statistical parameters: limits of agreement from Bland-Altman analysis, significance between differences (P value), and intraclass correlation coefficient (ICC).
RESULTS: Seventy-six eyes of 38 volunteers were measured. Interexaminer reproducibility for subjective refraction was excellent (ICC ≥0.96; P > .05), with low 95% confidence interval (CI) values for the power vectors M (spherical equivalent of the given refractive error), J0 (Jackson cross-cylinder, axes at 180 degrees and 90 degrees), and J45 (Jackson cross-cylinder, axes at 45 degrees and 135 degrees) (±0.51 diopter [D], ±0.14 D, and ±0.14 D, respectively). No significant differences in subjective refraction and visual acuity were found between the visual simulator and gold standard (P > .05), with 95% CIs for M, J0, and J45 (subjective refraction) of ±0.67 D, ±0.14 D, and ±0.16 D, respectively, and a ±0.10 logarithm of the minimum angle of resolution (visual acuity).
CONCLUSION: Subjective refraction results using the adaptive optics visual simulator agreed with those of the gold standard and can be used as the baseline for visual simulation of any optical corneal profile or intraocular lens design for refractive surgery patients.
SETTING: Optics Laboratory, University of Murcia, Murcia, Spain.
DESIGN: Prospective case series.
METHODS: Using the adaptive optics visual simulator, 2 examiners measured the subjective refraction and visual acuity in healthy eyes of volunteers; 1 examiner also used a trial frame as a gold standard. The interexaminer reproducibility and agreement with the gold standard were estimated using the following statistical parameters: limits of agreement from Bland-Altman analysis, significance between differences (P value), and intraclass correlation coefficient (ICC).
RESULTS: Seventy-six eyes of 38 volunteers were measured. Interexaminer reproducibility for subjective refraction was excellent (ICC ≥0.96; P > .05), with low 95% confidence interval (CI) values for the power vectors M (spherical equivalent of the given refractive error), J0 (Jackson cross-cylinder, axes at 180 degrees and 90 degrees), and J45 (Jackson cross-cylinder, axes at 45 degrees and 135 degrees) (±0.51 diopter [D], ±0.14 D, and ±0.14 D, respectively). No significant differences in subjective refraction and visual acuity were found between the visual simulator and gold standard (P > .05), with 95% CIs for M, J0, and J45 (subjective refraction) of ±0.67 D, ±0.14 D, and ±0.16 D, respectively, and a ±0.10 logarithm of the minimum angle of resolution (visual acuity).
CONCLUSION: Subjective refraction results using the adaptive optics visual simulator agreed with those of the gold standard and can be used as the baseline for visual simulation of any optical corneal profile or intraocular lens design for refractive surgery patients.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app