Assessment of material loss of retrieved magnetically controlled implants for limb lengthening.
Proceedings of the Institution of Mechanical Engineers. Part H, Journal of Engineering in Medicine 2018 November
PURPOSE: We aimed to understand wear from the telescopic component of PRECICE nails, which are used for distraction osteogenesis of the femur or tibia. We also aimed to identify any correlation between implant performance and patient factors.
METHODS: This retrieval study involved 11 magnetically controlled intramedullary nails from nine patients who had achieved the targeted leg length. All the nails were assessed macroscopically and microscopically for wear. All implants were radiographed to assess the internal mechanism. A Talyrond 365 (Taylor Hobson, Leicester, UK) roundness measuring machine was used to generate three-dimensional surface maps of the telescopic components to allow for measurement of wear.
RESULTS: Visual assessment of all the nails showed evidence of wear from the telescopic component. The radiographs revealed that all the nails had intact internal mechanism and no evidence of fractured pins. The roundness measuring machine showed that the quantity of wear was lowest in the latest design of the PRECICE nail. There was no significant correlation between wear and the two patient factors (duration of the lengthening phase, the time of implantation) included in this study.
CONCLUSION: This study is the first to investigate the performance of the three different designs of the PRECICE system with a focus on wear. We found that the latest design had the best implant performance. We are confident of the continued success of the PRECICE system and reassure surgeons and patients that they are unlikely to encounter problems with the implant related to wear.
METHODS: This retrieval study involved 11 magnetically controlled intramedullary nails from nine patients who had achieved the targeted leg length. All the nails were assessed macroscopically and microscopically for wear. All implants were radiographed to assess the internal mechanism. A Talyrond 365 (Taylor Hobson, Leicester, UK) roundness measuring machine was used to generate three-dimensional surface maps of the telescopic components to allow for measurement of wear.
RESULTS: Visual assessment of all the nails showed evidence of wear from the telescopic component. The radiographs revealed that all the nails had intact internal mechanism and no evidence of fractured pins. The roundness measuring machine showed that the quantity of wear was lowest in the latest design of the PRECICE nail. There was no significant correlation between wear and the two patient factors (duration of the lengthening phase, the time of implantation) included in this study.
CONCLUSION: This study is the first to investigate the performance of the three different designs of the PRECICE system with a focus on wear. We found that the latest design had the best implant performance. We are confident of the continued success of the PRECICE system and reassure surgeons and patients that they are unlikely to encounter problems with the implant related to wear.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app