Real-time dual-modal vein imaging system.
PURPOSE: In this paper, we present a vein imaging system to combine reflectance mode visible spectrum images (VIS) with transmission mode near-infrared (NIR) images in real time. Clear vessel localization is achieved in this manner with combined NIR-VIS dual-modal imaging.
METHODS: Transmission and reflectance mode optical instrumentation is used to combine VIS and NIR images. Two methods of displaying the combined images are demonstrated here. We have conducted experiments to determine the system's resolution, alignment accuracy, and depth penetration. Vein counts were taken from the hands of test subjects using the system and compared with vein counts taken by visual analysis.
RESULTS: Results indicate that the system can improve vein detection in the human hand while detecting veins of a diameter < 0.5 mm at any working distance and of a 0.25 mm diameter at an optimal working distance of about 30 cm. The system has also been demonstrated to clearly detect silicone vessels with artificial blood of diameter 2, 1, and 0.5 mm diameter under a tissue depth of 3 mm. In a study involving 25 human subjects, we have demonstrated that vein visibility was significantly increased using our system.
CONCLUSIONS: The results indicate that the device is a high-resolution solution to near-surface venous imaging. This technology can be applied for IV placement, morphological analysis for disease state detection, and biometric analysis.
METHODS: Transmission and reflectance mode optical instrumentation is used to combine VIS and NIR images. Two methods of displaying the combined images are demonstrated here. We have conducted experiments to determine the system's resolution, alignment accuracy, and depth penetration. Vein counts were taken from the hands of test subjects using the system and compared with vein counts taken by visual analysis.
RESULTS: Results indicate that the system can improve vein detection in the human hand while detecting veins of a diameter < 0.5 mm at any working distance and of a 0.25 mm diameter at an optimal working distance of about 30 cm. The system has also been demonstrated to clearly detect silicone vessels with artificial blood of diameter 2, 1, and 0.5 mm diameter under a tissue depth of 3 mm. In a study involving 25 human subjects, we have demonstrated that vein visibility was significantly increased using our system.
CONCLUSIONS: The results indicate that the device is a high-resolution solution to near-surface venous imaging. This technology can be applied for IV placement, morphological analysis for disease state detection, and biometric analysis.
Full text links
Related Resources
Trending Papers
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app