Using Microphone Technology to Improve Speech Perception in Noise in Children with Cochlear Implants.
Journal of the American Academy of Audiology 2018 October
BACKGROUND: Cochlear implant (CI) users are affected more than their normal hearing (NH) peers by the negative consequences of background noise on speech understanding. Research has shown that adult CI users can improve their speech recognition in challenging listening environments by using dual-microphone beamformers, such as adaptive directional microphones (ADMs) and wireless remote microphones (RMs). The suitability of these microphone technologies for use in children with CIs is not well-understood nor widely accepted.
PURPOSE: To assess the benefit of ADM or RM technology on speech perception in background noise in children and adolescents with cochlear implants (CIs) with no previous or current use of ADM or RM.
RESEARCH DESIGN: Mixed, repeated measures design.
STUDY SAMPLE: Twenty (20) children, ten (10) CI users (mean age 14.3 yrs) who used Advanced Bionics HiRes90K implants with research Naida processors, and ten (10) NH age-matched controls participated in this prospective study.
INTERVENTION: CI users listened with an ear-canal level microphone, T-Mic (TM), an ADM, and a wireless RM at different audio-mixing ratios. Speech understanding with five microphone settings (TM 100%, ADM, RM + TM 50/50, RM + TM 75/25, RM 100%) was evaluated in quiet and in noise.
DATA COLLECTION AND ANALYSIS: Speech perception ability was measured using children's spondee words to obtain a speech recognition threshold for 80% accuracy (SRT80%) in 20-talker babble where the listener sat in a sound booth 1 m (3.28') from the target speech (front) and noise (behind) to test five microphone settings (TM 100%, ADM, RM + TM 50/50, RM + TM 75/25, RM 100%). Group performance-intensity functions were computed for each listening condition to show the effects of microphone configuration with respect to signal-to-noise ratio (SNR). A difference score (CI Group minus NH Group) was computed to show the effect of microphone technology at different SNRs relative to NH. Statistical analysis using a repeated-measures analysis of variance evaluated the effects of the microphone configurations on SRT80% and performance at SNRs. Between-groups analysis of variance was used to compare the CI group with the NH group.
RESULTS: The speech recognition was significantly poorer for children with CI than children with NH in quiet and in noise when using the TM alone. Adding the ADM or RM provided a significant improvement in speech recognition for the CI group over use of the TM alone in noise (mean dB advantage ranged from 5.8 for ADM to 16 for RM100). When children with CI used the RM75 or RM100 in background babble, speech recognition was not statistically different from the group with NH.
CONCLUSION: Speech recognition in noise performance improved with the use of ADM and RM100 or RM75 over TM-only for children with CIs. Alhough children with CI remain at a disadvantage as compared with NH children in quiet and more favorable SNRs, microphone technology can enhance performance for some children with CI to match that of NH peers in contexts with negative SNRs.
PURPOSE: To assess the benefit of ADM or RM technology on speech perception in background noise in children and adolescents with cochlear implants (CIs) with no previous or current use of ADM or RM.
RESEARCH DESIGN: Mixed, repeated measures design.
STUDY SAMPLE: Twenty (20) children, ten (10) CI users (mean age 14.3 yrs) who used Advanced Bionics HiRes90K implants with research Naida processors, and ten (10) NH age-matched controls participated in this prospective study.
INTERVENTION: CI users listened with an ear-canal level microphone, T-Mic (TM), an ADM, and a wireless RM at different audio-mixing ratios. Speech understanding with five microphone settings (TM 100%, ADM, RM + TM 50/50, RM + TM 75/25, RM 100%) was evaluated in quiet and in noise.
DATA COLLECTION AND ANALYSIS: Speech perception ability was measured using children's spondee words to obtain a speech recognition threshold for 80% accuracy (SRT80%) in 20-talker babble where the listener sat in a sound booth 1 m (3.28') from the target speech (front) and noise (behind) to test five microphone settings (TM 100%, ADM, RM + TM 50/50, RM + TM 75/25, RM 100%). Group performance-intensity functions were computed for each listening condition to show the effects of microphone configuration with respect to signal-to-noise ratio (SNR). A difference score (CI Group minus NH Group) was computed to show the effect of microphone technology at different SNRs relative to NH. Statistical analysis using a repeated-measures analysis of variance evaluated the effects of the microphone configurations on SRT80% and performance at SNRs. Between-groups analysis of variance was used to compare the CI group with the NH group.
RESULTS: The speech recognition was significantly poorer for children with CI than children with NH in quiet and in noise when using the TM alone. Adding the ADM or RM provided a significant improvement in speech recognition for the CI group over use of the TM alone in noise (mean dB advantage ranged from 5.8 for ADM to 16 for RM100). When children with CI used the RM75 or RM100 in background babble, speech recognition was not statistically different from the group with NH.
CONCLUSION: Speech recognition in noise performance improved with the use of ADM and RM100 or RM75 over TM-only for children with CIs. Alhough children with CI remain at a disadvantage as compared with NH children in quiet and more favorable SNRs, microphone technology can enhance performance for some children with CI to match that of NH peers in contexts with negative SNRs.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
Perioperative echocardiographic strain analysis: what anesthesiologists should know.Canadian Journal of Anaesthesia 2024 April 11
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app