A Simple and Accurate Measurement Procedure for Serum Chloride by Ion Chromatography and Bias Evaluation of Six Ion Selective Electrode Measurements in China.
Clinical Laboratory 2018 September 2
BACKGROUND: Chloride is the main anion in human body. A simple and accurate method for serum chloride measurement was developed using ion chromatography (IC) in China.
METHODS: In the measurement, serum samples were diluted 500 times with water, filtered and injected into ion chromatography column. A mixed eluent (2 mmol/L CO32- + 12 mmol/L OH-) was used and peak area signal was collected. Five calibrators made from Standard Reference Material (SRM) 919b were used in the bracketing method. The IC method was applied as the comparative method and six ion selective electrode (ISE) measurement systems were evaluated using 60 individual patient serums.
RESULTS: The IC method was proven to be accurate. The precision was 0.18% - 0.30%, the recovery was 99.66% - 100.60%, the bias was 0.19% to -0.06%, and the related expanded uncertainty was 0.775% (k = 2). The precisions of the ISE systems were smaller than the 0.9% tolerable CV except for the Beckman DXC (0.91% - 1.16%). In comparison, the results of linear regression analysis showed that the correlation coefficients were 0.9876 to 0.9979. For all systems, the range of mean biases was -5.96 - 1.48 mmol/L (-5.57% - 1.36%); the expected biases at the medical decision levels were -4.97% - 0.84% at 90 mmol/L and -6.02% - 1.76% at 120 mmol/L. All biases of the Beckman AU met the requirement of within ± 1.5%.
CONCLUSIONS: The IC measurement method is proven to be of high precision and trueness, and the quality of routine ISE measurement of serum chloride still needs significant improvement. The establishment of the IC method can improve the measurement quality and promote its standardization process in China.
METHODS: In the measurement, serum samples were diluted 500 times with water, filtered and injected into ion chromatography column. A mixed eluent (2 mmol/L CO32- + 12 mmol/L OH-) was used and peak area signal was collected. Five calibrators made from Standard Reference Material (SRM) 919b were used in the bracketing method. The IC method was applied as the comparative method and six ion selective electrode (ISE) measurement systems were evaluated using 60 individual patient serums.
RESULTS: The IC method was proven to be accurate. The precision was 0.18% - 0.30%, the recovery was 99.66% - 100.60%, the bias was 0.19% to -0.06%, and the related expanded uncertainty was 0.775% (k = 2). The precisions of the ISE systems were smaller than the 0.9% tolerable CV except for the Beckman DXC (0.91% - 1.16%). In comparison, the results of linear regression analysis showed that the correlation coefficients were 0.9876 to 0.9979. For all systems, the range of mean biases was -5.96 - 1.48 mmol/L (-5.57% - 1.36%); the expected biases at the medical decision levels were -4.97% - 0.84% at 90 mmol/L and -6.02% - 1.76% at 120 mmol/L. All biases of the Beckman AU met the requirement of within ± 1.5%.
CONCLUSIONS: The IC measurement method is proven to be of high precision and trueness, and the quality of routine ISE measurement of serum chloride still needs significant improvement. The establishment of the IC method can improve the measurement quality and promote its standardization process in China.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app