Clinical commissioning of a new patient positioning system, SyncTraX FX4, for intracranial stereotactic radiotherapy.
Journal of Applied Clinical Medical Physics 2018 November
BACKGROUND & AIMS: A new real-time tracking radiotherapy (RTRT) system, the SyncTraX FX4 (Shimadzu, Kyoto, Japan), consisting of four X-ray tubes and four ceiling-mounted flat panel detectors (FPDs) combined with a linear accelerator, was installed at Uonuma Kikan Hospital (Niigata, Japan) for the first time worldwide. In addition to RTRT, the SyncTraX FX4 system enables bony structure-based patient verification. Here we provide the first report of this system's clinical commissioning for intracranial stereotactic radiotherapy (SRT).
MATERIALS & METHODS: A total of five tests were performed for the commissioning: evaluations of (1) the system's image quality; (2) the imaging and treatment coordinate coincidence; and (3) the localization accuracy of cone-beam computed tomography (CBCT) and SyncTraX FX4; (4) the measurement of air kerma; (5) an end-to-end test.
RESULTS & DISCUSSION: The tests revealed the following. (1) All image quality evaluation items satisfied each acceptable criterion in all FPDs. (2) The maximum offsets among the centers were ≤0.40 mm in all combinations of the FPD and X-ray tubes (preset). (3) The isocenter localization discrepancies between CBCT and preset #3 in the SyncTraX FX4 system were 0.29 ± 0.084 mm for anterior-posterior, -0.19 ± 0.13 mm for superior-inferior, 0.076 ± 0.11 mm for left-right, -0.11 ± 0.066° for rotation, -0.14 ± 0.064° for pitch, and 0.072±0.058° for roll direction. the Pearson's product-moment correlation coefficient between the two systems was >0.98 in all directions. (4) The mean air kerma value for preset #3 was 0.11 ± 0.0002 mGy in predefined settings (80 kV, 200 mA, 50 msec). (5) For 16 combinations of gantry and couch angles, median offset value in all presets was 0.31 mm (range 0.14-0.57 mm).
CONCLUSION: Our results demonstrate a competent performance of the SyncTraX FX4 system in terms of the localization accuracy for intracranial SRT.
MATERIALS & METHODS: A total of five tests were performed for the commissioning: evaluations of (1) the system's image quality; (2) the imaging and treatment coordinate coincidence; and (3) the localization accuracy of cone-beam computed tomography (CBCT) and SyncTraX FX4; (4) the measurement of air kerma; (5) an end-to-end test.
RESULTS & DISCUSSION: The tests revealed the following. (1) All image quality evaluation items satisfied each acceptable criterion in all FPDs. (2) The maximum offsets among the centers were ≤0.40 mm in all combinations of the FPD and X-ray tubes (preset). (3) The isocenter localization discrepancies between CBCT and preset #3 in the SyncTraX FX4 system were 0.29 ± 0.084 mm for anterior-posterior, -0.19 ± 0.13 mm for superior-inferior, 0.076 ± 0.11 mm for left-right, -0.11 ± 0.066° for rotation, -0.14 ± 0.064° for pitch, and 0.072±0.058° for roll direction. the Pearson's product-moment correlation coefficient between the two systems was >0.98 in all directions. (4) The mean air kerma value for preset #3 was 0.11 ± 0.0002 mGy in predefined settings (80 kV, 200 mA, 50 msec). (5) For 16 combinations of gantry and couch angles, median offset value in all presets was 0.31 mm (range 0.14-0.57 mm).
CONCLUSION: Our results demonstrate a competent performance of the SyncTraX FX4 system in terms of the localization accuracy for intracranial SRT.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app