Evaluation of Simulated Lesions as Surrogates to Clinical Lesions for Thoracic CT Volumetry: The Results of an International Challenge.
Academic Radiology 2018 September 13
RATIONALE AND OBJECTIVES: To evaluate a new approach to establish compliance of segmentation tools with the computed tomography volumetry profile of the Quantitative Imaging Biomarker Alliance (QIBA); and determine the statistical exchangeability between real and simulated lesions through an international challenge.
MATERIALS AND METHODS: The study used an anthropomorphic phantom with 16 embedded physical lesions and 30 patient cases from the Reference Image Database to Evaluate Therapy Response with pathologically confirmed malignancies. Hybrid datasets were generated by virtually inserting simulated lesions corresponding to physical lesions into the phantom datasets using one projection-domain-based method (Method 1), two image-domain insertion methods (Methods 2 and 3), and simulated lesions corresponding to real lesions into the Reference Image Database to Evaluate Therapy Response dataset (using Method 2). The volumes of the real and simulated lesions were compared based on bias (measured mean volume differences between physical and virtually inserted lesions in phantoms as quantified by segmentation algorithms), repeatability, reproducibility, equivalence (phantom phase), and overall QIBA compliance (phantom and clinical phase).
RESULTS: For phantom phase, three of eight groups were fully QIBA compliant, and one was marginally compliant. For compliant groups, the estimated biases were -1.8 ± 1.4%, -2.5 ± 1.1%, -3 ± 1%, -1.8 ± 1.5% (±95% confidence interval). No virtual insertion method showed statistical equivalence to physical insertion in bias equivalence testing using Schuirmann's two one-sided test (±5% equivalence margin). Differences in repeatability and reproducibility across physical and simulated lesions were largely comparable (0.1%-16% and 7%-18% differences, respectively). For clinical phase, 7 of 16 groups were QIBA compliant.
CONCLUSION: Hybrid datasets yielded conclusions similar to real computed tomography datasets where phantom QIBA compliant was also compliant for hybrid datasets. Some groups deemed compliant for simulated methods, not for physical lesion measurements. The magnitude of this difference was small (<5.4%). While technical performance is not equivalent, they correlate, such that, volumetrically simulated lesions could potentially serve as practical proxies.
MATERIALS AND METHODS: The study used an anthropomorphic phantom with 16 embedded physical lesions and 30 patient cases from the Reference Image Database to Evaluate Therapy Response with pathologically confirmed malignancies. Hybrid datasets were generated by virtually inserting simulated lesions corresponding to physical lesions into the phantom datasets using one projection-domain-based method (Method 1), two image-domain insertion methods (Methods 2 and 3), and simulated lesions corresponding to real lesions into the Reference Image Database to Evaluate Therapy Response dataset (using Method 2). The volumes of the real and simulated lesions were compared based on bias (measured mean volume differences between physical and virtually inserted lesions in phantoms as quantified by segmentation algorithms), repeatability, reproducibility, equivalence (phantom phase), and overall QIBA compliance (phantom and clinical phase).
RESULTS: For phantom phase, three of eight groups were fully QIBA compliant, and one was marginally compliant. For compliant groups, the estimated biases were -1.8 ± 1.4%, -2.5 ± 1.1%, -3 ± 1%, -1.8 ± 1.5% (±95% confidence interval). No virtual insertion method showed statistical equivalence to physical insertion in bias equivalence testing using Schuirmann's two one-sided test (±5% equivalence margin). Differences in repeatability and reproducibility across physical and simulated lesions were largely comparable (0.1%-16% and 7%-18% differences, respectively). For clinical phase, 7 of 16 groups were QIBA compliant.
CONCLUSION: Hybrid datasets yielded conclusions similar to real computed tomography datasets where phantom QIBA compliant was also compliant for hybrid datasets. Some groups deemed compliant for simulated methods, not for physical lesion measurements. The magnitude of this difference was small (<5.4%). While technical performance is not equivalent, they correlate, such that, volumetrically simulated lesions could potentially serve as practical proxies.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app