LMA® ProtectorTM Airway. First experience with a new second generation laryngeal mask

Fabian Eckardt, Jörg Engel, Sacha Tw Mann, Matthias Müller, Thomas Zajonz, Christian M Koerner, Michael Sander, Valesco Mann
Minerva Anestesiologica 2018 September 10
30207132

BACKGROUND: The LMA ProtectorTM Airway (LMA-Protector) is a new supraglottic airway promising a better seal, an improved drainage of gastric secretions and the opportunity of a simplified fiberscopic guided tracheal intubation. The aim of this study was to present a primary evaluation of the LMA-Protector in a clinical setting.

METHODS: After informed consent 50 patients, scheduled for minor/moderate surgery in supine position, were recruited. Pharyngeal seal pressures were examined in neutral position of the patients ́ head and in maximum passive extension of the neck. Additionally the fiberscopic view on the glottis was graduated and the feasibility of fiberscope guided tracheal intubation through the device was evaluated.

RESULTS: The median pharyngeal seal pressure of the LMA-Protector in neutral position of the head was 34 cmH2O. Passive extension of the neck did not cause a reduction of the pharyngeal seal (median pharyngeal seal pressure: 34.7 cmH2O; p< 0.039). The LMA-Protector was applicable for fiberscopic tracheal intubation, but is not reliable for blind tracheal intubation.

CONCLUSIONS: The LMA-Protector provides a high pharyngeal seal. Uncommon for laryngeal masks its pharyngeal seal is not affected by the extension of the patient ́s neck. As a second generation supraglottic airway which is also suitable for simplified fiberscopic guided tracheal intubation, the LMA-Protector could be considered as a supraglottic airway of the third generation.

Comments

You need to log in or sign up for an account to be able to comment.

No comments yet, be the first to post one!

Trending on Read

Available on the App Store

Available on the Play Store
Remove bar
Read by QxMD icon Read
30207132
×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"