3D ultrashort echo time MRI of the lung using stack-of-spirals and spherical k-Space coverages: Evaluation in healthy volunteers and parenchymal diseases.

BACKGROUND: Ultrashort echo time (UTE) has been shown to improve lung MRI quality in three dimensions. The evaluation of 3D-UTE stack-of-spirals VIBE (3D-USV) sequence for parenchymal diseases and a comparison of performance with that of a spherical mode of acquisition is needed.

PURPOSE: To assess MRI quality using a prototypical 3D-USV sequence and to compare performance with that of a spherical acquisition using Pointwise Encoding Time Reduction with Radial Acquisition (PETRA).

STUDY TYPE: Monocenter, prospective.

POPULATION: Twelve healthy volunteers and 32 adult patients with either cystic fibrosis (CF; n = 16) or interstitial lung disease (ILD; n = 16).

FIELD STRENGTH/SEQUENCE: Both free-breathing 3D-USV and PETRA were completed at 1.5T.

ASSESSMENT: In healthy volunteers, visual analysis of imaging quality was scored using a Likert scale. Quantitative evaluation of apparent signal ratio (Sr) and contrast ratio (Cr) was measured. Patients with CF and ILD completed both computed tomography (CT) and MRI. Depiction of structural alterations was assessed using dedicated clinical scores. All evaluations were done in consensus by two readers.

STATISTICAL TESTS: Comparison of means was assessed using the Wilcoxon signed rank test. Concordance and agreement between CT and MRI were assessed using the intraclass correlation coefficient (ICC) and kappa test.

RESULTS: In controls, 3D-USV yielded lower artifacts owing to better automatic respiratory synchronization than PETRA (P < 0.001). However, Sr and Cr of 3D-USV were found significantly lower by 2.25- and 2.36-fold, respectively (P < 0.001). In patients, 3D-USV and PETRA showed comparable performances to assess airway severity in CF (Bhalla score, ICC = 0.89 and ICC = 0.92, respectively) and presence of structural alterations in ILD such as honeycombing (kappa = 0.68 and kappa = 0.69, respectively).

DATA CONCLUSION: 3D-USV enables high-resolution morphological imaging of the lung without need of an external device to compensate respiratory motions. Automation and robustness of the method may facilitate clinical application for both airway and interstitial lung investigations.

LEVEL OF EVIDENCE: 2 Technical Efficacy: Stage 1 J. Magn. Reson. Imaging 2018.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app