Low-cost optical scanner and 3-dimensional printing technology to create lead shielding for radiation therapy of facial skin cancer: First clinical case series.

Purpose: Three-dimensional printing has been implemented at our institution to create customized treatment accessories, including lead shields used during radiation therapy for facial skin cancer. To effectively use 3-dimensional printing, the topography of the patient must first be acquired. We evaluated a low-cost, structured-light, 3-dimensional, optical scanner to assess the clinical viability of this technology.

Methods and materials: For ease of use, the scanner was mounted to a simple gantry that guided its motion and maintained an optimum distance between the scanner and the object. To characterize the spatial accuracy of the scanner, we used a geometric phantom and an anthropomorphic head phantom. The geometric phantom was machined from plastic and included hemispherical and tetrahedral protrusions that were roughly the dimensions of an average forehead and nose, respectively. Polygon meshes acquired by the optical scanner were compared with meshes generated from high-resolution computed tomography images. Most optical scans contained minor artifacts. Using an algorithm that calculated the distances between the 2 meshes, we found that most of the optical scanner measurements agreed with those from the computed tomography scanner within approximately 1 mm for the geometric phantom and approximately 2 mm for the head phantom. We used this optical scanner along with 3-dimensional printer technology to create custom lead shields for 10 patients receiving orthovoltage treatments of nonmelanoma skin cancers of the face. Patient, tumor, and treatment data were documented.

Results: Lead shields created using this approach were accurate, fitting the contours of each patient's face. This process added to patient convenience and addressed potential claustrophobia and medical inability to lie supine.

Conclusions: The scanner was found to be clinically acceptable, and we suggest that the use of an optical scanner and 3-dimensional printer technology become the new standard of care to generate lead shielding for orthovoltage radiation therapy of nonmelanoma facial skin cancer.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app