Quantitative comparison of a laterally misplaced pedicle screw with a re-directed screw. How much pull-out strength is lost?
Acta Orthopaedica et Traumatologica Turcica 2018 September 4
OBJECTIVE: Redirecting of a laterally misplaced pedicle screw into the accurate position decreases the pull-out strength due to the reinsertion, lateral wall cortical perforation and widening of the pedicle hole. Thus, this biomechanical study was performed to quantitatively analyze the pullout strength of a redirected laterally misplaced pedicle screw into the accurate position.
METHODS: Thirty pedicules of 15 bovine vertebrae were separated to 3 groups, according to the screw placement method: 1) standard flawless trajectory; 2) trajectory with lateral pedicle wall perforation; 3) trajectory with lateral wall perforation redirected to the standard trajectory. Samples were placed on a universal testing machine and pullout loads were measured. Kruskal-Wallis test was utilized within 95% confidence interval and p value <0.05 to test for the statistical significance.
RESULTS: The mean pullout strength was 2891±654,2 N(1383-3814,5) in Group 1; 817,8±227,6 N(308,6-1144,9) in Group 2 and 2081,1±487,7 N(1583,5-2962,5) in Group 3. The results found out to be statistically significant (p<0.05). Inter-group comparisons revealed that lateral pedicle wall perforation significantly decreases the pullout strength (p<0.05) and redirection of the screw increases the strength (p<0.05), however it was still weaker than the screws with flawless standard trajectory but this was not statistically significant (p>0.05).
CONCLUSION: The results of this study confirm that pullout strength of pedicle screw decreases by approximately 71% when the lateral wall is perforated and decreases 28% after redirection to the accurate position.
METHODS: Thirty pedicules of 15 bovine vertebrae were separated to 3 groups, according to the screw placement method: 1) standard flawless trajectory; 2) trajectory with lateral pedicle wall perforation; 3) trajectory with lateral wall perforation redirected to the standard trajectory. Samples were placed on a universal testing machine and pullout loads were measured. Kruskal-Wallis test was utilized within 95% confidence interval and p value <0.05 to test for the statistical significance.
RESULTS: The mean pullout strength was 2891±654,2 N(1383-3814,5) in Group 1; 817,8±227,6 N(308,6-1144,9) in Group 2 and 2081,1±487,7 N(1583,5-2962,5) in Group 3. The results found out to be statistically significant (p<0.05). Inter-group comparisons revealed that lateral pedicle wall perforation significantly decreases the pullout strength (p<0.05) and redirection of the screw increases the strength (p<0.05), however it was still weaker than the screws with flawless standard trajectory but this was not statistically significant (p>0.05).
CONCLUSION: The results of this study confirm that pullout strength of pedicle screw decreases by approximately 71% when the lateral wall is perforated and decreases 28% after redirection to the accurate position.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app