Microcatheter navigation through the clot: does size matter?

BACKGROUND: Despite high recanalization rates achieved with endovascular treatment of acute ischemic strokes, around 50% of eligible patients will not achieve a good outcome. Parameters that may determine patient outcomes include: time from puncture to recanalization, the collateral status, the anesthesia regimen, blood pressure management, and distal emboli. Characterization of distal emboli generated during mechanical thrombectomy has been performed in previous studies.

OBJECTIVE: To further investigate the risk of distal embolization associated with microcatheter navigation across the clot.

METHODS: A contrast-enhanced clot analog was used in an in vitro model that mimicked a middle cerebral artery occlusion within a complete circle of Willis vascular replica. The clot was crossed with one of the following microcatheters: Pro18, XT-27 or 3MAX. The emboli generated during the procedure were collected and measured.

RESULTS: The use of Pro18 and XT-27 resulted in a significant reduction of visible particles (size ≥500 µm) as compared with the 3MAX catheter (P<0.003). For the size range between 8 and 200 µm, there was a trend for Pro18 to generate fewer particles (-18%) than XT-27 but without statistical significance (P>0.05). In comparison with previously published data, acquired under the same conditions, it was found that the clot crossing maneuver accounts approximately for 12% of the total number of small emboli (<200 µm) induced during a stent retriever-mediated mechanical thrombectomy procedure via a balloon guide catheter.

CONCLUSIONS: The clot crossing maneuver has a significant effect on the total number of small particles induced during mechanical thrombectomy. Smaller microcatheter sizes should be favored when possible.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app