Fluoroscopic balloon diameter measurement at different pressures during Eustachian balloon dilation.

OBJECTIVES: To measure the diameter of inflated balloons at different pressures during Eustachian tube (ET) balloon dilation under fluoroscopic guidance.

DESIGN: Prospective cohort study.

SETTING: Tertiary academic referral centre.

PARTICIPANTS: Eighteen patients who underwent ET balloon dilation with use of a balloon catheter, 20 mm long and 6 mm in diameter, under combined endoscopic and fluoroscopic guidance.

MAIN OUTCOME MEASURES: Degrees of inflation at three different portions (proximal, middle and distal) of the balloon at controlled pressures (3, 5, 8 and 10 atmospheres [atm]) and at the maximum pressure manually applied.

RESULTS: The mean proximal, middle and distal diameters of the inflated balloons were 5.3 ± 0.4 mm, 5.3 ± 0.4 mm and 4.9 ± 0.5 mm at 10 atm. The distal diameters were significantly smaller than middle and proximal diameters at all the pressures (P < 0.01). When compared to the distal diameter (100%, 4.9 ± 0.5 mm) at 10 atm, the distal diameters were 73% (3.6 ± 0.6 mm) at 3 atm, 88% (4.3 ± 0.5 mm) at 5 atm and 96% (4.7 ± 0.4 mm) at 8 atm. The distal diameter (4.1 ± 0.3 mm) at the maximum pressure manually applied was in between those at 3 and 5 atm.

CONCLUSIONS: The distal diameter of the balloon increased significantly as a function of the pressure and most (88%) inflation occurred at a low pressure of 5 atm, which was sufficient to inflate the distal diameter of the balloon more than 3 mm. The manual pressurisation could inflate a balloon by as much as could be expected, at between 3 and 5 atm.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app