Validation of frame-based positioning accuracy with cone-beam computed tomography in Gamma Knife Icon radiosurgery.
Physica Medica : PM 2018 August
BACKGROUND: Frame based positioning accuracy in Gamma Knife (GK) stereotactic radiosurgery (SRS) is extremely high but removal of a post may be necessary to enable the treatment in selected patients.
OBJECTIVE: To verify the positioning accuracy in clinical scenarios with 4 and 3 posts in patients and phantoms using cone-beam CT (CBCT) of Gamma Knife Icon™.
METHODS: We analyzed positioning accuracy for 12 patients with standard 4 post setup using pretreatment CBCT (pre-CBCT) on GK Icon™ and report 4 patients with different clinical scenarios (removal of a post). We performed phantom measurements to verify the frame accuracy via CBCT in different clinical scenarios without the influence of the human patient.
RESULTS: Mean frame accuracy for 12 patients with 4 posts was 0.35 mm/0.34 degree. Mean motion during treatment was 0.11 mm/0.04 degree. For two of the clinical scenarios where a post was removed, we found acceptable deviations within 0.66 mm/0.61 degree. For 2 patients, a deviation of 2.94 mm/-3.47 degree and 1.85 mm/-0.74 degree was found and replanning was necessary. Phantom measurements showed good agreement when planning MR/CT was performed with 4 or 3 post. Larger deviations of 0.86 mm/0.88 degree were detected when a post was removed after planning MR/CT.
CONCLUSION: The frame accuracy with 4 posts before and during GK treatments is as high as expected. For clinical situations, where a post is removed after planning-CT/MR, pre-treatment position verification is strongly suggested using stereotactic CBCT or the P-CT/MR should be repeated to avoid possible mistreatments.
OBJECTIVE: To verify the positioning accuracy in clinical scenarios with 4 and 3 posts in patients and phantoms using cone-beam CT (CBCT) of Gamma Knife Icon™.
METHODS: We analyzed positioning accuracy for 12 patients with standard 4 post setup using pretreatment CBCT (pre-CBCT) on GK Icon™ and report 4 patients with different clinical scenarios (removal of a post). We performed phantom measurements to verify the frame accuracy via CBCT in different clinical scenarios without the influence of the human patient.
RESULTS: Mean frame accuracy for 12 patients with 4 posts was 0.35 mm/0.34 degree. Mean motion during treatment was 0.11 mm/0.04 degree. For two of the clinical scenarios where a post was removed, we found acceptable deviations within 0.66 mm/0.61 degree. For 2 patients, a deviation of 2.94 mm/-3.47 degree and 1.85 mm/-0.74 degree was found and replanning was necessary. Phantom measurements showed good agreement when planning MR/CT was performed with 4 or 3 post. Larger deviations of 0.86 mm/0.88 degree were detected when a post was removed after planning MR/CT.
CONCLUSION: The frame accuracy with 4 posts before and during GK treatments is as high as expected. For clinical situations, where a post is removed after planning-CT/MR, pre-treatment position verification is strongly suggested using stereotactic CBCT or the P-CT/MR should be repeated to avoid possible mistreatments.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app