A beam model for focused proton pencil beams.
Physica Medica : PM 2018 August
INTRODUCTION: We present a beam model for Monte Carlo simulations of the IBA pencil beam scanning dedicated nozzle installed at the Skandion Clinic. Within the nozzle, apart from entrance and exit windows and the two ion chambers, the beam traverses vacuum, allowing for a beam that is convergent downstream of the nozzle exit.
MATERIALS AND METHODS: We model the angular, spatial and energy distributions of the beam phase space at the nozzle exit with single Gaussians, controlled by seven energy dependent parameters. The parameters were determined from measured profiles and depth dose distributions. Verification of the beam model was done by comparing measured and GATE acquired relative dose distributions, using plan specific log files from the machine to specify beam spot positions and energy.
RESULTS: GATE-based simulations with the acquired beam model could accurately reproduce the measured data. The gamma index analysis comparing simulated and measured dose distributions resulted in >95% global gamma index pass rates (3%/2 mm) for all depths.
CONCLUSION: The developed beam model was found to be sufficiently accurate for use with GATE e.g. for applications in quality assurance (QA) or patient motion studies with the IBA pencil beam scanning dedicated nozzles.
MATERIALS AND METHODS: We model the angular, spatial and energy distributions of the beam phase space at the nozzle exit with single Gaussians, controlled by seven energy dependent parameters. The parameters were determined from measured profiles and depth dose distributions. Verification of the beam model was done by comparing measured and GATE acquired relative dose distributions, using plan specific log files from the machine to specify beam spot positions and energy.
RESULTS: GATE-based simulations with the acquired beam model could accurately reproduce the measured data. The gamma index analysis comparing simulated and measured dose distributions resulted in >95% global gamma index pass rates (3%/2 mm) for all depths.
CONCLUSION: The developed beam model was found to be sufficiently accurate for use with GATE e.g. for applications in quality assurance (QA) or patient motion studies with the IBA pencil beam scanning dedicated nozzles.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app