Design Analysis and Optimization of a Single-Layer PDMS Microfluidic Artificial Lung.
IEEE Transactions on Bio-medical Engineering 2019 April
OBJECTIVE: Microfluidic artificial lungs (μALs) are being researched for future clinical use due to the potential for increased gas exchange efficiency, small blood contacting surface area, small priming volume, and biomimetic blood flow paths. However, a current roadblock to clinical use is the need to stack hundreds to thousands of these small-scale μALs in parallel to reach clinically relevant blood flows. The need for so many layers not only increases the complexity and projected cost to manufacture a μAL, but also could result in devices which are cumbersome, and, therefore, not suitable for portable artificial lung systems.
METHODS: Here, we describe the design analysis and optimization of a single-layer μAL that simultaneously calculates rated blood flow, blood contacting surface area, blood volume, pressure drop, and shear stress as a function of blood channel height using previously developed closed-form mathematical equations. A μAL designed using this procedure is then implemented and tested.
RESULTS: The resulting device exhibits a rated flow of 17 mL/min and reduces the number of layers required for clinically relevant μAL devices by a factor of up to 32X compared to previous work.
CONCLUSION: This procedure could significantly reduce manufacturing complexity as well as eliminate a barrier to the clinical application of these promising devices.
SIGNIFICANCE: The described method results in the highest rated flow for any single-layer μAL to date.
METHODS: Here, we describe the design analysis and optimization of a single-layer μAL that simultaneously calculates rated blood flow, blood contacting surface area, blood volume, pressure drop, and shear stress as a function of blood channel height using previously developed closed-form mathematical equations. A μAL designed using this procedure is then implemented and tested.
RESULTS: The resulting device exhibits a rated flow of 17 mL/min and reduces the number of layers required for clinically relevant μAL devices by a factor of up to 32X compared to previous work.
CONCLUSION: This procedure could significantly reduce manufacturing complexity as well as eliminate a barrier to the clinical application of these promising devices.
SIGNIFICANCE: The described method results in the highest rated flow for any single-layer μAL to date.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app