Study on the Accuracy of Structural and FSI Heart Valves Simulations.
Cardiovascular Engineering and Technology 2018 December
PURPOSE: The performance of heart valves, either native or artificial, can be evaluated by means of finite element analyses, either from a structural or a fluid-structure interaction (FSI) point of view. The latter captures the coupling between the valve leaflets and the blood in a more realistic way. The selection of the appropriate finite elements approach for the model is the first and fundamental step to achieve accurate simulations. The aim of this work is to investigate the influence of the type, formulation, size, and shape of the elements in heart valves simulations.
METHODS: The effects related to the choice of the finite elements-shell or solid- in structural and FSI simulations were analyzed. In particular, the analysis of grid convergence on both the structure and fluid domains, the influence of the element typology, formulation and damping factor in an idealized three-leaflets valve model loaded with physiological pressure conditions were investigated.
RESULTS: Stress values and valve kinematics results confirmed the importance of performing a proper verification process for selecting the most appropriate elements with the optimal accuracy to computational cost ratio.
CONCLUSION: In this regard, our results indicate the quadrangular shell with reduced integration and viscous hourglass control to be the best choice to model heart valves. If a solid discretization is required, quadratic hexahedral elements with full integration are also acceptable. Finally, our results show that the damping coefficient needs to be carefully selected in order to smooth out the high frequency modes of the structure without introducing excessive numerical artificial viscosity.
METHODS: The effects related to the choice of the finite elements-shell or solid- in structural and FSI simulations were analyzed. In particular, the analysis of grid convergence on both the structure and fluid domains, the influence of the element typology, formulation and damping factor in an idealized three-leaflets valve model loaded with physiological pressure conditions were investigated.
RESULTS: Stress values and valve kinematics results confirmed the importance of performing a proper verification process for selecting the most appropriate elements with the optimal accuracy to computational cost ratio.
CONCLUSION: In this regard, our results indicate the quadrangular shell with reduced integration and viscous hourglass control to be the best choice to model heart valves. If a solid discretization is required, quadratic hexahedral elements with full integration are also acceptable. Finally, our results show that the damping coefficient needs to be carefully selected in order to smooth out the high frequency modes of the structure without introducing excessive numerical artificial viscosity.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app