Comparative Evaluation of Accuracy of Two Electronic Apex Locators in the Presence of Contemporary Irrigants: An In vitro Study.
Aim: The present study was aimed to compare the accuracy of Root ZX Mini and Propex II in the presence of 0.1% octinidine dohydrochloride (OCT), 2% chlorhexidine gluconate (CHX), and 5% sodium hypochlorite (NaOCl) heated and nonheated before and after preparation.
Materials and Methods: Eighty extracted single-rooted teeth were selected for the study and decoronated. Teeth were mounted in an alginate model. Actual working length (AL) was measured using a stereomicroscope under ×4 magnification. Electronic working length measurements were recorded using Root ZX Mini and Propex II apex locators in the presence of 0.1% OCT, 2% CHX, and 5% NaOCl (nonheated and heated to 60°C) before and after preparation. Mean and standard deviation differences before and after preparation were calculated and statistically analyzed using analysis of variance and paired t -test.
Results: The accuracy of Root ZX Mini before and after preparation within ±0.5 mm of AL was consistently high in the presence of irrigants than Propex II. 5% NaOCl (heated and nonheated) showed more variation than the other irrigants, in the working length determination in both the apex locators.
Conclusion: Electronic length measurements were shorter with heated and nonheated 5% NaOCl and longer with 0.1% OCT and 2% CHX for both the electronic apex locators.
Materials and Methods: Eighty extracted single-rooted teeth were selected for the study and decoronated. Teeth were mounted in an alginate model. Actual working length (AL) was measured using a stereomicroscope under ×4 magnification. Electronic working length measurements were recorded using Root ZX Mini and Propex II apex locators in the presence of 0.1% OCT, 2% CHX, and 5% NaOCl (nonheated and heated to 60°C) before and after preparation. Mean and standard deviation differences before and after preparation were calculated and statistically analyzed using analysis of variance and paired t -test.
Results: The accuracy of Root ZX Mini before and after preparation within ±0.5 mm of AL was consistently high in the presence of irrigants than Propex II. 5% NaOCl (heated and nonheated) showed more variation than the other irrigants, in the working length determination in both the apex locators.
Conclusion: Electronic length measurements were shorter with heated and nonheated 5% NaOCl and longer with 0.1% OCT and 2% CHX for both the electronic apex locators.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app