Improving accuracy of corneal power measurement with partial coherence interferometry after corneal refractive surgery using a multivariate polynomial approach.
Biomedical Engineering Online 2018 August 14
BACKGROUND: To improve accuracy of IOLMaster (Carl Zeiss, Jena, Germany) in corneal power measurement after myopic excimer corneal refractive surgery (MECRS) using multivariate polynomial analysis (MPA).
METHODS: One eye of each of 403 patients (mean age 31.53 ± 8.47 years) was subjected to MECRS for a myopic defect, measured as spherical equivalent, ranging from - 9.50 to - 1 D (mean - 4.55 ± 2.20 D). Each patient underwent a complete eye examination and IOLMaster scan before surgery and at 1, 3 and 6 months follow up. Axial length (AL), flatter keratometry value (K1), steeper keratometry value (K2), mean keratometry value (KM) and anterior chamber depth measured from the corneal endothelium to the anterior surface of the lens (ACD) were used in a MPA to devise a method to improve accuracy of KM measurements.
RESULTS: Using AL, K1, K2 and ACD measured after surgery in polynomial degree 2 analysis, mean error of corneal power evaluation after MECRS was + 0.16 ± 0.19 D.
CONCLUSIONS: MPA was found to be an effective tool in devising a method to improve precision in corneal power evaluation in eyes previously subjected to MECRS, according to our results.
METHODS: One eye of each of 403 patients (mean age 31.53 ± 8.47 years) was subjected to MECRS for a myopic defect, measured as spherical equivalent, ranging from - 9.50 to - 1 D (mean - 4.55 ± 2.20 D). Each patient underwent a complete eye examination and IOLMaster scan before surgery and at 1, 3 and 6 months follow up. Axial length (AL), flatter keratometry value (K1), steeper keratometry value (K2), mean keratometry value (KM) and anterior chamber depth measured from the corneal endothelium to the anterior surface of the lens (ACD) were used in a MPA to devise a method to improve accuracy of KM measurements.
RESULTS: Using AL, K1, K2 and ACD measured after surgery in polynomial degree 2 analysis, mean error of corneal power evaluation after MECRS was + 0.16 ± 0.19 D.
CONCLUSIONS: MPA was found to be an effective tool in devising a method to improve precision in corneal power evaluation in eyes previously subjected to MECRS, according to our results.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app