Reduction thermal damage to cortical bone using ultrasonically-assisted drilling.
Technology and Health Care : Official Journal of the European Society for Engineering and Medicine 2018
BACKGROUND: Thermal damage induced by bone drilling is a common problem during surgical procedures. A recent and promising method utilizes high-frequency low-amplitude vibration in the feed direction during drilling and has the potential to reduce drilling temperature, minimizing the risk of thermal damage.
OBJECTIVE: The purpose of this study was to investigate the effects of ultrasonically-assisted drilling (UAD) on cortical bone temperature.
METHODS: A series of experiments was conducted to compare the cortical bone temperature during UAD with that during conventional drilling (CD). A thermo-mechanical 3D finite element model (FEM) of UAD was developed, using ABAQUS, to help understand temperature changes during drilling of cortical bone. The numerical simulation results of FEM showed good agreement with the experimental data. Subsequently, a predictive model was developed for bone temperature during drilling, using multiple regression analysis based on the results from numerical simulation.
RESULTS: The results showed drill diameter had the greatest influence on drilling temperature, followed by the rotational speed of the drill. Additionally, the variation of vibration frequency had more influence on the drilling temperature than did the amplitude.
CONCLUSIONS: Ultrasonically-assisted drilling is helpful to lower drilling temperature to reduce the thermal damage of bone tissue.
OBJECTIVE: The purpose of this study was to investigate the effects of ultrasonically-assisted drilling (UAD) on cortical bone temperature.
METHODS: A series of experiments was conducted to compare the cortical bone temperature during UAD with that during conventional drilling (CD). A thermo-mechanical 3D finite element model (FEM) of UAD was developed, using ABAQUS, to help understand temperature changes during drilling of cortical bone. The numerical simulation results of FEM showed good agreement with the experimental data. Subsequently, a predictive model was developed for bone temperature during drilling, using multiple regression analysis based on the results from numerical simulation.
RESULTS: The results showed drill diameter had the greatest influence on drilling temperature, followed by the rotational speed of the drill. Additionally, the variation of vibration frequency had more influence on the drilling temperature than did the amplitude.
CONCLUSIONS: Ultrasonically-assisted drilling is helpful to lower drilling temperature to reduce the thermal damage of bone tissue.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app