Smoothing Effect in Vital Sign Recordings: Fact or Fiction? A Retrospective Cohort Analysis of Manual and Continuous Vital Sign Measurements to Assess Data Smoothing in Postoperative Care.

BACKGROUND: Data smoothing of vital signs has been reported in the anesthesia literature, suggesting that clinical staff are biased toward measurements of normal physiology. However, these findings may be partially explained by clinicians interpolating spurious values from noisy signals and by the undersampling of physiological changes by infrequent manual observations. We explored the phenomenon of data smoothing using a method robust to these effects in a large postoperative dataset including respiratory rate, heart rate, and oxygen saturation (SpO2). We also assessed whether the presence of the vital sign taker creates an arousal effect.

METHODS: Study data came from a UK upper gastrointestinal postoperative ward (May 2009 to December 2013). We compared manually recorded vital sign data with contemporaneous continuous data recorded from monitoring equipment. We proposed that data smoothing increases differences between vital sign sources as vital sign abnormality increases. The primary assessment method was a mixed-effects model relating continuous-manual differences to vital sign values, adjusting for repeated measurements. We tested the regression slope significance and predicted the continuous-manual difference at clinically important vital sign values. We calculated limits of agreement (LoA) between vital sign sources using the Bland-Altman method, adjusting for repeated measures. Similarly, we assessed whether the vital sign taker affected vital signs, comparing continuous data before and during manual recording.

RESULTS: From 407 study patients, 271 had contemporaneous continuous and manual recordings, allowing 3740 respiratory rate, 3844 heart rate, and 3896 SpO2 paired measurements for analysis. For the model relating continuous-manual differences to continuous-manual average vital sign values, the regression slope (95% confidence interval) was 0.04 (-0.01 to 0.10; P = .11) for respiratory rate, 0.04 (-0.01 to 0.09; P = .11) for heart rate, and 0.10 (0.07-0.14; P < .001) for SpO2. For SpO2 measurements of 91%, the model predicted a continuous-manual difference (95% confidence interval) of -0.88% (-1.17% to -0.60%). The bias (LoA) between measurement sources was -0.74 (-7.80 to 6.32) breaths/min for respiratory rate, -1.13 (-17.4 to 15.1) beats/min for heart rate, and -0.25% (-3.35% to 2.84%) for SpO2. The bias (LoA) between continuous data before and during manual observation was -0.57 (-5.63 to 4.48) breaths/min for respiratory rate, -0.71 (-10.2 to 8.73) beats/min for heart rate, and -0.07% (-2.67% to 2.54%) for SpO2.

CONCLUSIONS: We found no evidence of data smoothing for heart rate and respiratory rate measurements. Although an effect exists for SpO2 measurements, it was not clinically significant. The wide LoAs between continuous and manually recorded vital signs would commonly result in different early warning scores, impacting clinical care. There was no evidence of an arousal effect caused by the vital sign taker.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app