Sparse Detector Configuration in SiPM Digital Photon Counting PET: a Feasibility Study.
Molecular Imaging and Biology : MIB : the Official Publication of the Academy of Molecular Imaging 2019 June
PURPOSE: To investigate the minimum number of SiPM detectors required for solid-state digital photon counting (DPC) oncologic whole-body 2-deoxy-2-[18 F]fluoro-D-glucose ([18 F]FDG) positron emission tomography (PET)/X-ray computed tomography (CT).
PROCEDURES: A DPC PET/CT (Vereos, Philips) with 23,040 1-to-1 crystal-to-detector couplings was utilized. [18 F]FDG PET/CT of a uniformity phantom and 10 oncology patients selected by block randomization from a large clinical trial were included (457 ± 38 MBq, 64 ± 22 min p.i, body mass index (BMI) of 14-41). Sparse-ring PET configurations with 50 % detector reduction in tangential and axial directions were analyzed and compared to the current full ring configuration. Resulting images were reviewed blindly and quantitatively over detectable lesions and the liver.
RESULTS: One hundred twelve lesions (d = 10 to 95 mm) were analyzed in the patient population. All lesions remained visible and were demonstrated without compromised image quality under all BMIs in the 50 % sparse detector configurations despite the DPC PET system sensitivity reduction to 1/4th. An excellent consistency of SUVmax measurements of lesions with an average of 5 % SUVmax difference was found between dPET of full and sparse configurations.
CONCLUSIONS: The feasibility of either expanding the axial field of view (FOV) by a factor of two or halving the number of detectors was demonstrated for solid-state digital photon counting PET, thus either potentially enabling cost reduction or extended effective axial FOV without increased cost.
PROCEDURES: A DPC PET/CT (Vereos, Philips) with 23,040 1-to-1 crystal-to-detector couplings was utilized. [18 F]FDG PET/CT of a uniformity phantom and 10 oncology patients selected by block randomization from a large clinical trial were included (457 ± 38 MBq, 64 ± 22 min p.i, body mass index (BMI) of 14-41). Sparse-ring PET configurations with 50 % detector reduction in tangential and axial directions were analyzed and compared to the current full ring configuration. Resulting images were reviewed blindly and quantitatively over detectable lesions and the liver.
RESULTS: One hundred twelve lesions (d = 10 to 95 mm) were analyzed in the patient population. All lesions remained visible and were demonstrated without compromised image quality under all BMIs in the 50 % sparse detector configurations despite the DPC PET system sensitivity reduction to 1/4th. An excellent consistency of SUVmax measurements of lesions with an average of 5 % SUVmax difference was found between dPET of full and sparse configurations.
CONCLUSIONS: The feasibility of either expanding the axial field of view (FOV) by a factor of two or halving the number of detectors was demonstrated for solid-state digital photon counting PET, thus either potentially enabling cost reduction or extended effective axial FOV without increased cost.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app